Chapter 3: Installation
AirLink 7032 Gas Monitor Operator’s Manual
11. Clamp down the enclosure latches.
Fault Indicator Connection
The Fault terminal provides an output to power some form of Fault indicator. The Fault terminal
is wet contact, uses the same DC voltage that is powering the device, provides 500 mA maximum,
and is a DC only output.
The fault terminal’s failsafe operation can be configured as described on page 33.
1. Open the enclosure box to expose the Front Panel.
2. Unscrew the two thumb-screws on the Front Panel.
3. Open the Front Panel so that the Terminal Board is exposed.
4. Locate the Fault Terminal Block on the terminal board.
5. Connect a positive (red) wire to the terminal labeled “+”.
6. Connect a negative (black) wire to the terminal labeled “-”.
Figure 8: Fault Indicator Wiring