65-2670XL-XX-07 M2A-XL Transmitter
Chapter 7: Maintenance
•
47
Replacing the Detector
NOTE:
In most cases, it is only necessary to replace the plug-in sensor.
1. Turn off or unplug power to the M2A-XL.
2. Remove the detector junction box cover.
3. Disconnect the detector leads from the terminal block in the detector junction box. Note the
position of the color-coded leads as you remove them.
4. Unscrew the detector from the detector junction box.
5. Guide the detector leads of the replacement detector through the bottom conduit hub of the
detector junction box, then screw the mounting threads of the detector into the conduit hub. If
necessary for environmental conditions, apply thread sealant or Teflon tape to the hub and/or
detector threads to seal them.
6. Connect the detector leads to the terminal block the same way the old detector was wired (see
7. Reinstall the detector junction box cover.
8. Turn on or plug in power to the M2A-XL.
CAUTION:
For all gases except ClO
2
, allow the replacement detector to warm up
for 15 minutes before you continue with the next step.
For ClO
2
, allow the replacement detector to warm up for 2-3 hours
before you continue with the next step.
9. Calibrate the replacement detector as described in “Calibration” on page 41.