42
•
Chapter 7: Maintenance
65-2640XL-04 M2A Transmitter Operator’s Manual
Adjusting the Fresh Air Reading
1. While in normal operation, press and hold the UP/YES button for 5 seconds to enter
Calibration Mode. Release the button when the following screen appears.
2. If you want to continue with calibration, press and release the UP/YES button. The display
will indicate the target gas and
CAL Mode
for a few seconds before showing
FreshAir
Adjust?
.
If you want to exit Calibration Mode, press and release the DOWN/NO button. The M2A
will indicate
Leaving CAL Mode
and the M2A will return to normal operation.
3. To continue with adjusting the fresh air reading, press and release the UP/YES button.
ENTER
will alternate with
FreshAir
on the top display line and the current gas reading will
be on the bottom display line.
To skip adjusting the fresh air reading, press and release the DOWN/NO button. The
display will indicate
SPAN w/Cal Gas?
. Skip to the next section, Adjusting the Span
4. Screw the zero air cylinder into the fixed flow regulator.
5. Use the flexible tubing coming from the calibration adapter/splash guard to connect the
regulator to the calibration adapter.
6. Turn the regulator knob counterclockwise to open the regulator.
7. Allow zero air to flow for the length of time determined in “Determining Response Time” on
8. Turn the regulator knob clockwise to close the regulator.
9. Unscrew the zero air cylinder from the fixed flow regulator. Leave the calibration adapter/
splash guard’s flexible tubing connected to the regulator. The M2A will continue to display
the minimum gas response on the display and retain the response level in its memory.
10. Press and release the ENTER button. The M2A will perform a fresh air adjustment and the
display will indicate
SPAN w/Cal Gas?
.
NOTE:
If the fresh air adjustment fails, see “Troubleshooting” on page 37 for
recommended actions.
Adjusting the Span Setting
1. If you want to continue with adjusting the span setting, press and release the UP/YES button.
APPLY
will alternate with
SPAN Gas
on the top display line and the current gas reading will
be on the bottom display line.
If you want to skip adjusting the span reading, press and release the DOWN/NO button.
The display will indicate
Leaving Cal Mode
and the M2A will return to normal operation.
Calib?
YES/NO