65-2640XL-04 M2A Transmitter Operator’s Manual
Chapter 3: Installation & Startup
•
19
Figure 7: Wiring the Detector to the M2A
14. Reinstall the detector terminal strip into its socket.
15. If shielded cable is used between the detector junction box and the M2A junction box, connect
the cable’s drain wire to an available chassis (earth) ground at the M2A junction box. Do not
connect the shield drain wire at the detector junction box.
Black
Green
White
Red
Detector
Terminal
Strip
C
NC
N
O
AL
AR
M
1
C
NC
N
O
AL
AR
M
2
C
N
C
NO
FA
IL
Cable
Shield
M2A-XL
J-Box
Install conduit seal
within 18 inches
Green
Calibration Adapter/Splash Guard
Combustible Gas
Detector
Detector J-Box
Install conduit seal within
2 inches (for Group B
environments) or 18
inches (for Groups C and
D environments)
3/4 NPT Conduit Opening
Black
White
Red
-
S
+
A
B
C
RS
4
8
5
PW
R
/SI
G
R
W
G
B
LE
L