background image

 

18 • 65-2396RK CO

2

 Transmitter

4. Use the following formula to determine the correct test points output for the calibrating sample.

Output (mV) = (calibrating sample/fullscale) X 400 + 100

For example, with a calibrating sample of 2.5 %Vol. and a fullscale setting of 5 %Vol., the correct 
output is 300 mV.

300(mV) = (50/100) X 400 +100

5. Screw the calibration cup onto the detector housing.
6. Place the controller into its calibration mode or disable external alarms.

NOTE:

Calibrating the CO

2

 transmitter may cause alarms. Be sure to put the controller into its 

calibration mode or disable external alarms before continuing.

Setting the Zero Reading

Since there is a background of CO

2

 in air of typically 300 - 600 ppm (0.03 - 0.06% volume), it is 

necessary to use a calibration kit with a 100% nitrogen cylinder to set the zero signal of a CO

2

 

transmitter with a low range detector. Fresh air can be used to zero the transmitter if a 0-20% volume, 
0-50% volume, or 0-100% volume detector is being used.

1. Screw the regulator into the 100% nitrogen calibration cylinder.

2. Use the calibration kit sample tubing to connect the regulator to the calibration cup.
3. Turn the regulator’s on/off knob counterclockwise to open the regulator.
4. Allow the gas to flow for one minute and verify a reading of 100 mV (±2 mV). If necessary, use 

the zero pot on the amplifier to adjust the reading to 100 mV (±2 mV).

5. Turn the regulator’s on/off knob clockwise to close the regulator.
6. Unscrew the regulator from the 100% nitrogen calibration cylinder. 
7. Leave the sample tubing connected to the regulator and the calibration cup.

Setting the Response Reading

1. Screw the regulator into the calibration cylinder. Verify that the calibration gas is representative of 

the transmitter’s target gas.

2. Turn the regulator’s on/off knob counterclockwise to open the regulator.
3. Allow the calibration gas to flow for one minute and verify that the reading matches the response 

reading (±2 mV) you determined earlier. If necessary, use the span pot on the amplifier to adjust 
the reading to match the correct response reading.

4. Turn the regulator’s on/off knob clockwise to close the regulator.
5. Unscrew the regulator from the calibration cylinder.

Returning to Normal Operation

1. Remove the voltmeter leads from the amplifier test points.

2. Unscrew the calibration cup from the detector.

NOTE:

For convenience, leave the regulator and calibration cup connected by the sample tubing.

Содержание 65-2396RK

Страница 1: ...RKI Instruments Inc www rkiinstruments com 65 2396RK CO2 Transmitter Operator s Manual Part Number 71 0184RK Revision C Released 9 1 20...

Страница 2: ...alibration and maintenance of the transmitter is essential for proper operation and correct readings Please calibrate and maintain this transmitter regularly Frequency of calibration depends upon the...

Страница 3: ...nt costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRE...

Страница 4: ...ring the CO2 Transmitter 9 Startup 12 Introducing Incoming Power 12 Setting the Zero Signal 12 Maintenance 13 Preventive Maintenance 13 Troubleshooting 14 Replacing Components of the CO2 Transmitter 1...

Страница 5: ...nings in this manual to assure proper and safe operation of the 65 2396RK and to minimize the risk of personal injury Be sure to maintain and periodically calibrate the 65 2396RK as described in this...

Страница 6: ...that repels liquids is on one end of the detector and allows sample gas to enter the detector Four color coded leads red white green and black extend from the other end of the detector The leads allow...

Страница 7: ...ip See Wiring the CO2 Transmitter to a Controller on page 9 for all wiring procedures related to the transmitter Zero Pot The zero pot is located in the upper left corner of the amplifier see Figure 1...

Страница 8: ...ifier to the controller Use the cover on the front of the junction box to access the interior of the junction box The detector and amplifier are factory installed in the junction box Three spacers ins...

Страница 9: ...r facing down see Figure 2 Wiring the CO2 Transmitter to a Controller WARNING Always verify that the power to the controller is off before you make wiring connections 1 Turn off the controller 2 Turn...

Страница 10: ...e 3 Connect the positive wire to the PWR SIG terminal Connect the signal wire to the PWR SIG S terminal Connect the negative wire to the PWR SIG terminal CAUTION If shielded cable is used leave the ca...

Страница 11: ...e to an available chassis earth ground at the controller RKI controllers typically have a ground stud that can be used to ground the cable s drain wire 4 20 mA In S DC Ground 24 VDC Cable Shield Red W...

Страница 12: ...nce there is a background of CO2 in air of typically 300 600 ppm 0 03 0 06 volume it is necessary to use a calibration kit with a 100 nitrogen cylinder to set the zero signal of a CO2 transmitter with...

Страница 13: ...unction box Maintenance This section describes maintenance procedures It includes preventive maintenance troubleshooting and component replacement procedures Preventive Maintenance This section descri...

Страница 14: ...KI for further instruction Slow or No Response Difficult or Unable to Calibrate Unable to accurately set the zero or response reading during calibration Transmitter requires frequent calibration Note...

Страница 15: ...detector terminal strip as shown in Table 4 below and Figure 3 on page 11 of this manual 9 Reinstall the detector terminal strip into its socket 10 Reinstall the junction box cover 11 Turn on or plug...

Страница 16: ...the screw lock washer and flat washer you removed in Step 5 above 9 Install the detector and controller terminals strips into their sockets on the new amplifier as shown in Figure 3 on page 11 of thi...

Страница 17: ...manding for example detection in a clean temperature controlled environment where calibration adjustments are minimal at calibration then a calibration frequency of every 9 to 12 months is adequate If...

Страница 18: ...e calibration cup 3 Turn the regulator s on off knob counterclockwise to open the regulator 4 Allow the gas to flow for one minute and verify a reading of 100 mV 2 mV If necessary use the zero pot on...

Страница 19: ...ler display reading decreases and stabilizes at a typical background CO2 level The 0 5 000 ppm 0 9 000 ppm and 0 5 volume detectors will display a reading in a fresh air environment because they have...

Страница 20: ...ion box and amplifier 65 2396RK 03 Infrared CO2 transmitter 0 5 Volume includes detector junction box and amplifier 65 2396RK 04 Infrared CO2 transmitter 0 20 Volume includes detector junction box and...

Страница 21: ...ge and knob 0 5 liter minute continuous flow for 34 liter aluminum 58 liter 103 liter calibration cylinders cylinders with internal threads 81 1103RK Calibration cup 82 0006RK Pot adjust screwdriver f...

Отзывы: