background image

 

65-2396RK CO

2

 Transmitter • 17

11. Turn on the controller and place it into normal operation.

NOTE:

When first powered up, the transmitter will enter about a one minute  period when the 4-20 
mA output is stabilizing and may be above the controller alarm points or well below zero 
momentarily. RKI controllers have a one minute warmup period when the controller does not 
display any gas reading or give any alarm indication. The CO

2

 transmitter’s 4-20 mA signal 

should be stable by the time the controller’s warmup period is over. 

12. Allow the transmitter to warm-up for 5 minutes.
13. Calibrate the transmitter as described in “Calibration” on page 17 of this manual.

Calibration Frequency

Although there is no particular calibration frequency that is correct for all applications, a calibration 
frequency of every 6 months is adequate for most infrared CO

2

 transmitter applications. Unless 

experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a 
calibration frequency of every 6 months. 

If an application is not very demanding, for example detection in a clean, temperature controlled 
environment where calibration adjustments are minimal at calibration, then a calibration frequency of 
every 9 to 12 months is adequate. 

If an application is very demanding, for example if the environment is not well controlled, then more 
frequent calibration than every 6 months may be necessary. 

Calibration

This section describes how to calibrate the CO

2

 transmitter. It includes procedures to prepare for 

calibration, set the zero reading, set the response reading, and return to normal operation.

WARNING: The controller is not an active gas monitoring device during the calibration 

procedure.

NOTE:

The following procedure assumes the use of a calibration kit which includes a calibration gas 
cylinder, a 100% nitrogen cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a 
calibration cup for the detector, and a short piece of sample tubing to connect the regulator to 
the calibration cup.

Preparing for Calibration

1. Unscrew and remove the junction box cover.

2. Set a voltmeter to measure in the millivolt (mV) range.
3. Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the red (+) 

test point; plug the negative lead into the black (-) test point.

Содержание 65-2396RK

Страница 1: ...RKI Instruments Inc www rkiinstruments com 65 2396RK CO2 Transmitter Operator s Manual Part Number 71 0184RK Revision C Released 9 1 20...

Страница 2: ...alibration and maintenance of the transmitter is essential for proper operation and correct readings Please calibrate and maintain this transmitter regularly Frequency of calibration depends upon the...

Страница 3: ...nt costs local repair costs transportation costs or contingent expenses incurred without our prior approval THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS EXPRE...

Страница 4: ...ring the CO2 Transmitter 9 Startup 12 Introducing Incoming Power 12 Setting the Zero Signal 12 Maintenance 13 Preventive Maintenance 13 Troubleshooting 14 Replacing Components of the CO2 Transmitter 1...

Страница 5: ...nings in this manual to assure proper and safe operation of the 65 2396RK and to minimize the risk of personal injury Be sure to maintain and periodically calibrate the 65 2396RK as described in this...

Страница 6: ...that repels liquids is on one end of the detector and allows sample gas to enter the detector Four color coded leads red white green and black extend from the other end of the detector The leads allow...

Страница 7: ...ip See Wiring the CO2 Transmitter to a Controller on page 9 for all wiring procedures related to the transmitter Zero Pot The zero pot is located in the upper left corner of the amplifier see Figure 1...

Страница 8: ...ifier to the controller Use the cover on the front of the junction box to access the interior of the junction box The detector and amplifier are factory installed in the junction box Three spacers ins...

Страница 9: ...r facing down see Figure 2 Wiring the CO2 Transmitter to a Controller WARNING Always verify that the power to the controller is off before you make wiring connections 1 Turn off the controller 2 Turn...

Страница 10: ...e 3 Connect the positive wire to the PWR SIG terminal Connect the signal wire to the PWR SIG S terminal Connect the negative wire to the PWR SIG terminal CAUTION If shielded cable is used leave the ca...

Страница 11: ...e to an available chassis earth ground at the controller RKI controllers typically have a ground stud that can be used to ground the cable s drain wire 4 20 mA In S DC Ground 24 VDC Cable Shield Red W...

Страница 12: ...nce there is a background of CO2 in air of typically 300 600 ppm 0 03 0 06 volume it is necessary to use a calibration kit with a 100 nitrogen cylinder to set the zero signal of a CO2 transmitter with...

Страница 13: ...unction box Maintenance This section describes maintenance procedures It includes preventive maintenance troubleshooting and component replacement procedures Preventive Maintenance This section descri...

Страница 14: ...KI for further instruction Slow or No Response Difficult or Unable to Calibrate Unable to accurately set the zero or response reading during calibration Transmitter requires frequent calibration Note...

Страница 15: ...detector terminal strip as shown in Table 4 below and Figure 3 on page 11 of this manual 9 Reinstall the detector terminal strip into its socket 10 Reinstall the junction box cover 11 Turn on or plug...

Страница 16: ...the screw lock washer and flat washer you removed in Step 5 above 9 Install the detector and controller terminals strips into their sockets on the new amplifier as shown in Figure 3 on page 11 of thi...

Страница 17: ...manding for example detection in a clean temperature controlled environment where calibration adjustments are minimal at calibration then a calibration frequency of every 9 to 12 months is adequate If...

Страница 18: ...e calibration cup 3 Turn the regulator s on off knob counterclockwise to open the regulator 4 Allow the gas to flow for one minute and verify a reading of 100 mV 2 mV If necessary use the zero pot on...

Страница 19: ...ler display reading decreases and stabilizes at a typical background CO2 level The 0 5 000 ppm 0 9 000 ppm and 0 5 volume detectors will display a reading in a fresh air environment because they have...

Страница 20: ...ion box and amplifier 65 2396RK 03 Infrared CO2 transmitter 0 5 Volume includes detector junction box and amplifier 65 2396RK 04 Infrared CO2 transmitter 0 20 Volume includes detector junction box and...

Страница 21: ...ge and knob 0 5 liter minute continuous flow for 34 liter aluminum 58 liter 103 liter calibration cylinders cylinders with internal threads 81 1103RK Calibration cup 82 0006RK Pot adjust screwdriver f...

Отзывы: