RKC INSTRUMENT INC.

All Rights Reserved, Copyright 

 1995, RKC INSTRUMENT INC.

 

Resin Pressure Measuring System 

CZ-100P

 

(Resin Pressure Sensor)

 

PCT-300

 

(Output Converter) 

Instruction Manual

 

IM100CZ04-E4

 

Thank you for purchasing this RKC product. In order to achieve maximum 
performance and ensure proper operation of your new instrument, carefully 
read all the instructions in this manual. Please place the manual in a 
convenient location for easy reference. 

 

 

CONTENTS

 

1. OUTLINE ···························································   2 
2. PRODUCT CHECK ··············································   2 
3. MOUNTING ························································   2 
4. WIRING ·····························································   5 
5. ADJUSTMENT ····················································   6 
6. CALIBRATION ····················································   6 
7. LINEARIZING FUNCTION SETTING ·······················   8 
8. TROUBLES AND CAUSE······································   8 
9. EXPLANATION OF EACH TERMS··························   9 
10. SPECIFICATIONS ···········································   10 
 

SYMBOLS 

:  This mark indicates precautions that must be taken if 

there is danger of electric shock, fire, etc., which could 
result in loss of life or injury. 

:  This mark indicates that if these precautions and 

operating procedures are not taken, damage to the 
instrument may result. 

:  This mark indicates that all precautions should be taken 

for safe usage. 

: This mark indicates important information on 

installation, handling and operating procedures. 

: This mark indicates supplemental information on 

installation, handling and operating procedures. 

:  This mark indicates where additional information may 

be located. 

 

 

 

 

 To prevent injury to persons, damage to instrument and 

equipment, a suitable external protection device shall 
be required. 

 All wiring must be completed before power is turned on 

to prevent electric shock, fire or damage to instrument 
and equipment. 

 This instrument must be used in accordance with the 

specifications to prevent fire or damage to instrument 
and equipment.   

 This instrument is not intended for use in locations 

subject to flammable or explosive gases. 

 Do not touch high-voltage connections such as power 

supply terminals, etc. to avoid electric shock. 

 RKC is not responsible if this instrument is repaired, 

modified or disassembled by other than 
factory-approved personnel. Malfunction can occur and 
warranty is void under these conditions. 

 
 
 

 This product is intended for use with industrial machines, test and 

measuring equipment. (It is not designed for use with medical 
equipment and nuclear energy.) 

 This is a Class A instrument. In a domestic environment, this 

instrument may cause radio interference, in which case the user may 
be required to take additional measures. 

 This instrument is protected from electric shock by reinforced 

insulation. Provide reinforced insulation between the wire for the 
input signal and the wires for instrument power supply, source of 
power and loads. 

 Be sure to provide an appropriate surge control circuit respectively 

for the following: 
-  If input/output or signal lines within the building are longer than 30 

meters. 

-  If input/output or signal lines leave the building, regardless the 

length. 

 

 

This instrument is designed for installation in an enclosed 
instrumentation panel. All high-voltage connections such as power 
supply terminals must be enclosed in the instrumentation panel to 
avoid electric shock by operating personnel. 

 All precautions described in this manual should be taken to avoid 

damage to the instrument or equipment. 

  All wiring must be in accordance with local codes and regulations. 

 All wiring must be completed before power is turned on to prevent 

electric shock, instrument failure, or incorrect action. The power must be 
turned off before repairing work for input break and output failure 
including replacement of sensor, contactor or SSR, and all wiring must 
be completed before power is turned on again. 

 To prevent instrument damage as a result of failure, protect the power 

line and the input/output lines from high currents with a suitable 
overcurrent protection device with adequate breaking capacity such as 
fuse, circuit breaker, etc. 

 Prevent metal fragments or lead wire scraps from falling inside 

instrument case to avoid electric shock, fire or malfunction. 

 Tighten each terminal screw to the specified torque found in the 

manual to avoid electric shock, fire or malfunction. 

 For proper operation of this instrument, provide adequate ventilation for 

heat dispensation. 

 Do not connect wires to unused terminals as this will interfere with 

proper operation of the instrument. 

  Turn off the power supply before cleaning the instrument. 

  Do not use a volatile solvent such as paint thinner to clean the instrument. 

Deformation or discoloration will occur. Use a soft, dry cloth to remove 
stains from the instrument. 

 To avoid damage to instrument display, do not rub with an abrasive 

material or push front panel with a hard object. 

 The resin adhered to this instrument should be cleaned to dry cloth 

with a clean while the resin is still hardened, be careful of burns. 

 Tools such as wire wheels or abrasive cloths should never be used to 

clean the process diaphragm. 

 Do not apply impact to nor drop this product. If so, its damage or fault 

may result. 

 As precise parts are incorporated, do not give any shock and handle 

carefully during transportation and installation. Take great care not to 
scratch the diaphragm. 

 Do not wipe or rub the nameplate on the outer case of the resin 

pressure sensor with a cloth moistened with an organic solvent, or a 
glove. If so, the printed section may be erased. 

 This product uses stainless steel, aluminum, and fluororesin (O-rings 

and leadwire covering materials). When disposing of each part used 
for this product, always follows the procedure for disposing of industrial 
wastes stipulated by the respective local community. 

 If the output from the resin pressure sensor is found to be abnormal 

during the operation, immediately stop the operation and inspect for 
any distortion or damage of the diaphragm. If the diaphragm is 
damaged, the pressure of the measured media (e.g.resin) is applied to 
the inside of the resin pressure sensor. 
Continued use under such conditions may result in the damage of the 
screw which fixes the folder in the housing, and the folder in the 
housing may come off in the worst case. 

  Do not use the pressure sensor in stead of a blind bolt. 

 

NOTICE 

 This manual assumes that the reader has a fundamental knowledge of 

the principles of electricity, process control, computer technology and 
communications. 

  The figures, diagrams and numeric values used in this manual are only 

for purpose of illustration. 

 RKC is not responsible for any damage or injury that is caused as a 

result of using this instrument, instrument failure or indirect damage. 

 RKC is not responsible for any damage and/or injury resulting from the 

use of instruments made by imitating this instrument. 

 Periodic maintenance is required for safe and proper operation of this 

instrument. Some components have a limited service life, or 
characteristics that change over time. 

 Every effort has been made to ensure accuracy of all information 

contained herein. RKC makes no warranty expressed or implied, with 
respect to the accuracy of the information. The information in this 
manual is subject to change without prior notice. 

 No portion of this document may be reprinted, modified, copied, 

transmitted, digitized, stored, processed or retrieved through any 
mechanical, electronic, optical or other means without prior written 
approval from RKC. 

 

 

 

WARNING

 

CAUTION

 

!

 

CAUTION

 

!

 

WARNING 

Содержание CZ-100P

Страница 1: ...wer is turned on again To prevent instrument damage as a result of failure protect the power line and the input output lines from high currents with a suitable overcurrent protection device with adequ...

Страница 2: ...Diaphragm section material S SUS630 Standard H HASTELLOY C Optional 3 Diaphragm surface treatment N Standard K CERAMIC Kanigin plate Optional 4 Intrinsically safe N Standard For non explosionproof spe...

Страница 3: ...he slanting direction Fig 1 B to the horizontal direction Fig 1 C in this order In this case take measures of 1 and 2 if necessary by checking the sensor outer cylinder surface temperature To the rele...

Страница 4: ...llow loose fitting Z 1024 or Z 1024A type 0 to 1 MPa Z 1024 or Z 1024A type 24 0 0 2 3 1 5 Pressure sensor installation 1 Make sure the mounting hole is correctly machined If installing the pressure s...

Страница 5: ...maximum available current of 8 A Use the solderless terminal appropriate to the screw size Screw size M3 6 Recommended tightening torque 0 4 N m 4 kgf cm Specified solderless terminals With isolation...

Страница 6: ...de of PCT 300 refer to drawing on page 6 to 2 Gain setting for CZ 100P LLA Z 1024A type Set the RATED OUTPUT Rated output mV V value specified on the label of the pressure sensor at Gain setting in th...

Страница 7: ...wing correction equation and correction factor e1 e0 1 0 5 S K L 5 e0 Rated output of the resin pressure sensor e1 Rated output after the correction S Conductor nominal cross sectional area mm2 K Corr...

Страница 8: ...on is fluctuated Value different from the sensor rated output set to the rated output setter for the PCT 300 The diaphragm deteriorated deformed or damaged The sensor exposed to hot or cold wind Some...

Страница 9: ...aphragm deformed or damaged as the push rod pushes the diaphragm from the inside as a result of the deformation of the sensor flange by excessive tightening torque 9 EXPLANATION OF EACH TERMS Term Exp...

Страница 10: ...0 5 As the input side of bridge resistance the 350 5 type is also available This type is interchangeable with the 374 10 type Temperature characteristics Maximum diaphragm temperature 400 C Maximum st...

Страница 11: ...Linearity Within 0 01 of span Noise Within 0 1 p p of span 0 1 to 10 Hz Response 10 Hz 100 Hz transfer type Factory shipment 10 Hz Power supply 90 to 264 V AC Including power supply voltage variation...