User’s Manual
22 of26
While
pressing
the
pedal
controller , the system has no laser
exportation
Try to increase the work voltage to check whether the laser is
emits or not.
Check whether the system enters preparation mode or not.
Check whether the feet step switch is normal or not, if it is
damaged, replace it immediately.
Check whether the “work” key is pressed or not.
Please contact the dealer.
The system cannot start
Revolve the key to proper position.
The laser outputs to becomes
obviously weak or have" empty
gun", not a laser
The voltage might be too low and the equipment cannot
work normally.
The laser exportation lens is not clean; please wipe the lens
The laser light inside the machine is destroyed by fire,
replace the laser gun;
The instrument is too hot, shut down the laser for 30
minutes.
Screw the treatment head down, check the front mirror, if it
is damaged, replace immediately.
Examine the handle or a part of the treatment head whether
water leaks in, replace immediately if it leaks.
The laser treatment head is leaks
If the laser treatment head is damaged, please change the
laser treatment head or contact the dealer.
There is a ‘cheep’ sound in the
instrument
operation
process
when pressing preignition.
The instrument does not be used over a long period,
preignition is difficulty.
The indoor temperature is too low, adjust to rise the indoor
temperature properly, we suggest installing the air condition.
Check the circuit immediately.
The indoor humidity is too high, keep the room dry.
The light is damaged, replace the light.
The temperature of cooling
system is higher
Close the system; cool the system slightly, Re- start. Contact
the dealer.
The energy lowers gradually when
using
Check whether there are impurities on the mirror surface or
not. If there is dust clean it.
Whether the temperature of the handle is too hot or not,
Leave the machine to cool for 30 minutes or replace
circulating water.
Check if the front mirror is intact or not, if the mirror is
broken, please contact the seller.
The power of The laser treatment
head goes down
The life of laser treatment head is not long.
The laser breaks down
The light in the laser machine must be replaced. The
treatment head probably is damaged
,
please contact the
dealer.