User’s Manual
15 of26
15) Replace the treatment head or replace water
1. When replacing the treatment head, drain the remaining water in the previous handle. Pay
particular attention as not to spill any water on the electronic terminals. Wipe off any spilled water
with a soft absorbent cloth.
2. Replace water every 2-3 months, please carry on according to the following steps strictly.
Disconnect handpiece. But you need to keep the above part (as Figure 4.4)
on the water conduit of handpiece.
Turn on the infusion/drain tap and
unscrew
the ventilation screw caps. (as Figure 4.1)
To drain excess water from the cooling tank. Turn on the drain tap, insert the drainage hose
supplied then open the ventilation screw cap. Fasten all screw caps firmly.
1.
Only distilled water is to be used in the cooling system.
2.
When using the system for the first time, rinse the cooling system with distilled water.
3.
Ensure that the water level is adequate before treatment to prevent over heating.
4.
Distilled water is added periodically specially when working in warmer climates and when the
treatment handpiece are changed often.
5.
Make sure that the treatment head is firmly secured and the screw caps
are firmly screwed into place prior to operation.
3.4 Move
If the device is to be moved within the general area,
1. Turn the device off
2. Put the treatment head on the support
3. Pull out the power cable
4. Relocate the unit (Avoid tilting the device)
If the instrument is to be moved outside the hospital, please contact your dealer.
3.5 Transport and Storage Procedure
3.5.1 Procedure
1. The cooling water in the instrument should be let out and cleaned during storage and transport.
2. The device should vertically be installed in the outer packing box during transport. The
instrument is fixed on the base >550mm×490mm×430mm so that the wheels of the instrument
will not glide.