The No. 1 rotor and No. 2 rotor are designed for making measurements
using a JIS 300 mL beaker containing about 350 mL of sample fl uid. If
another container is used, the viscosity resistance working on the rotor
will be different, leading to deviations in the measurement result. If a
different container has to be used, make sure that it is considerably larger
than the JIS 300 mL beaker. In this case, the obtained measurement values
will be low.
The No. 3 rotor must be combined with the No. 3 cup, otherwise large
errors will occur in the reading.
*The actual capacity of the JIS 300 mL beaker is about 420 mL.