66
Serial Interface
CMS n1 n2 n3 n4 n5 n6 n7
Make comparator function settings
n1 to n2
Comparator level 00 to 98 (steps of 2)
n3
Delay time 0 to 9
n4=0
Auto reset OFF
n4=1
Auto reset ON
n5 to n6
Auto reset time 00 to 90 (steps of 10)
n7=0 Buzzer
OFF
n7=1 Buzzer
ON
For items that are not to be changed, send _ (space).
CMS?
Get comparator function settings
Output data format
p1 p2 p3 p4 p5 p6 p7<
EOT
><
CR
><
LF
>
p1 to p2
Comparator level 00 to 98
p3
0 to 9 (delay time)
p4
0 to 1 (auto reset ON/OFF)
p5 to p6
00 to 90 (auto reset time)
p7
0 to 1 (buzzer ON/OFF)