36
Measurement
Peak hold
The peak hold function uses 51.2 kHz sampling on the vibration input signal
to determine the peak value.
Setting procedure
1. Use the MENU key to call up menu 3.
2. Use the setting keys to select the setting 3-1 (peak hold ON).
3. Return to the measurement screen.
4. When the peak hold function has been activated, the indication PEAK HOLD appears
on the display, and the indicated numeric value is the peak value measured up to that
point. When a higher peak occurs, the display is updated.
When the peak hold value is higher than the overload threshold, the indication PEAK
HOLD OVER is shown.
5. The HOLD RESET key can be used to reset the peak hold value at
any time.
Содержание VM-83
Страница 2: ......
Страница 6: ...iv...
Страница 10: ...viii...
Страница 17: ...7 Controls and Features Side panel...
Страница 80: ...70 Technical Information 3 Measurement range with servo accelerometer LS 10C Acceleration measurement range...
Страница 81: ...71 Technical Information Velocity measurement range...
Страница 82: ...72 Technical Information Displacement measurement range...
Страница 97: ...87 Specifications 3 Measurement range with servo accelerometer LS 10C Acceleration measurement range...
Страница 99: ...89 Specifications Displacement measurement range...
Страница 102: ...92 Specifications Unit mm Dimensional drawings...
Страница 106: ...This product is environment friendly It does not include toxic chemicals on our policy No 63760 18 07...