Serial Interface
HPF n
Set high-pass fi lter setting
Command action depends on INPUT SELECT switch
setting.
Piezoelectric accelerometer
PICKUP, PREAMP1, PREAMP2
Servo accelerometer
SERVO , SERVO
Piezoelectric
Servo
n=0
OFF
0.1
n=1
1
0.3
n=2
3
1
n=3 10
1
n=4
20
1
n=5
50
1
n=6*
user fi lter
user fi lter
* This is valid only when the user fi lter is set up
by using VM-83PB1 software (option).
With piezoelectric accelerometer and measurement
mode VEL or DISP, n = 0 setting is not allowed.