4-1. Address Setting Method 

 

Appendix 

 
 

4-1. Address Setting Method 

 
Explanation of the keys 

 MODE 

key 

 ALM2/

 key 

 ALM1/

 key 

 TEST/SET 

key 

 
 
 
Setting procedure 

  Move to a maintenance mode by pressing and holding MODE key. 

(Display : [1-1

ZERO] ) 

  Select display [1-8

M MODE] by pressing 

▲▼

 key, and press SET key. 

(Display : [

----/

M MODE] ) 

  Press and hold SET key again. 

(Display : [2-0

GAS TEST] ) 

  Select display [2-6

SETTING2 ] by pressing 

▲▼

 key, and press SET key 

(Display : [SE 0

ADDRESS] ) 

  Press SET key 

(Display : [XX

MAC1] ) 

  Set address by pressing 

▲▼

 key and press SET key. 

(Display : [SE 0

ADDRESS] ) 

  Press and hold MODE key to complete the maintenance mode. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

MODE

ALM2

ALM1

TEST

SET

   

   

   

   

Содержание RM-5000 Series

Страница 1: ...T1E 0740 Indicator Alarm Unit RM 5000 Series RS 485 Modbus RTU Manual for Communication Function...

Страница 2: ...sure to read manuals of the equipment related to the network system Limitation of Liability We do not take the responsibility for the result which is caused by the program of the equipment which commu...

Страница 3: ...000 Modbus Register Map 4 2 2 Operation List of Status Data Address40001 6 3 Communication Specification 7 3 1 RM 5000 series 7 3 2 Modbus Protocol 8 4 Appendix 9 4 1 Address Setting Method 9 4 2 Wire...

Страница 4: ...owing Modbus function is available 0x03 Read Holding Register Exceptional response In the following cases RM 5000 responds exceptional response of Modbus 1 In a case that unavailable function is selec...

Страница 5: ...valent 4000 40003 Gas Concentration Data 2 3 Signed integer Significant figures of gas concentration which is made to integer Actual Gas concentration is calculated by multiplying this integer by bit...

Страница 6: ...1 of address 40028 40027 L side Digit Signed integer L side digit in double W range Significant figures of L side digit which is made to integer Actual L side digit is calculated by multiplying this i...

Страница 7: ...oncentration becomes an integer which maintains the number of significant digits Examples If the gas concentration is 20 9 in full scale 25 0 209 If the gas concentration is 0 2 in full scale 50 0 2 I...

Страница 8: ...Duplex Synchronization method Asynchronous Connection form N Maximum connectable number 127 ea against 1 x master unit Protocol Modbus RTU Available Function 03 Read Holdind Register Transmission spee...

Страница 9: ...16 bits and sending side calculates and adds the calculated value to the message Receiving side must confirm that the error check value is same as the calculated value If it is not same the received...

Страница 10: ...ey Display 1 1 ZERO Select display 1 8 M MODE by pressing key and press SET key Display M MODE Press and hold SET key again Display 2 0 GAS TEST Select display 2 6 SETTING2 by pressing key and press S...

Страница 11: ......

Страница 12: ......

Страница 13: ......

Страница 14: ......

Отзывы: