10 Product Specifications
10-1. List of specifications
SD-1DGH-AS
- 58 -
10
Product Specifications
10-1. List of specifications
[SD-1DGH-AS] <ATEX/IECEx Specifications>
Detection principle
Semiconductor
Gas to be detected
Combustible/toxic gas
Applicable gas detector head
SD-1DGH
Concentration display
Seven-segment LED (four-digit)
Detection range
Depends on the gas to be detected
Display resolution
Depends on the gas to be detected
Detection method
Aspirator suction type
Sampling condition
Equivalent to atmospheric pressure (compressed air: 0.1 - 0.7 MPa, maximum use amount: 3
L/min <converted into atmospheric pressure>)
Suction flow rate
0.8
±
0.1 L/min
Alarm setpoint
Depends on the gas to be detected
Power display
POWER lamp on (green)
External output
Gas concentration signal/alarm contact (gas alarm or fault alarm, gas/fault common alarm)
Alarm accuracy
(under the same conditions)
Within ±25% of the alarm setpoint value (combustible gas)/Within 30% of the alarm setpoint
value (toxic gas)
Response time
(under the same conditions)
Within 30 or 60 seconds after providing the gas 1.6 times the alarm setpoint (depending on the
gas to be detected)
Gas alarm type
One-step alarm (H)
Gas alarm display
ALM lamp on (red)
Gas alarm pattern
Auto-reset
Fault alarm/self diagnosis
System abnormalities/sensor abnormalities
Fault alarm display
FAULT lamp on (yellow)/detail display
Fault alarm pattern
Auto-reset
Alarm contact
No-voltage contact 1a, de-energized (energized in response to an alarm) or energized
(de-energized in response to an alarm)
Contact capacity
250 VAC - 0.5 A/30 VDC - 0.5 A (resistant load)
Transmission system
3-wire type analog transmission (common cable for power and signal <power, signal, common>)
Transmission specifications
4 - 20 mA DC (linear/load resistance under 300
Ω
, resolution divided into 200)
Transmission cable
CVVS1.25 sq - 3-core or CVVS1.25 sq - 5-core (when the contact is used)
Transmission distance
Below 1.25 km
Power supply
24 VDC (17 – 26.4 VDC)
Power Consumption
Maximum 3.1 W
Cable connecting port
Pressure proof packing gland
(Cable gland<G3/4> or Adapter A<NPT1/2> or Adapter B<NPT3/4>)
Tube connecting port
Gas IN: Rc1/4 (O.D ø6-1t, half-union for Cu or SUS tube <Bs or SUS> supplied)
OUT: O.D ø8-1t, flare union for Cu or SUS tube (Bs or SUS)
Compressed air IN: O.D ø8-1t, flare union for Cu tube
Initial clear
Approx. 25 seconds
Operating temperatures
-20 - 60ºC (at a constant condition)
Operating humidities
Below 95% RH (Non-condensing)
Structure
Wall mounting sampling unit (Cu or SUS type)
Explosion-proof structure
Flame-proof enclosures
Explosion-proof class
II
2G Ex db
II
C T5 Gb (ATEX) / Ex db
II
C T5 Gb (IECEx)
Содержание PT2-169
Страница 1: ...PT2E 2525 Smart Transmitter Gas Detector Head SD 1DGH AS Operating Manual PT2E 170 PT2 170...
Страница 4: ......
Страница 8: ......
Страница 9: ......
Страница 10: ......
Страница 12: ......
Страница 15: ......
Страница 19: ......
Страница 20: ......
Страница 22: ......
Страница 23: ......
Страница 24: ......
Страница 28: ......
Страница 29: ......
Страница 32: ......
Страница 33: ......
Страница 34: ......
Страница 35: ......
Страница 38: ......
Страница 43: ......
Страница 45: ......
Страница 46: ......
Страница 47: ......
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......
Страница 52: ......
Страница 53: ......
Страница 60: ...10 Product Specifications 10 1 List of specifications SD 1DGH AS 60 Adapter A NPT1 2 Adapter B NPT3 4...
Страница 62: ...10 Product Specifications 10 1 List of specifications SD 1DGH AS 62 Outline Drawings...
Страница 65: ......