5 Maintenance
5-2. Other adjustments/Cleaning method
- 25 -
PLU-70
5-2. Other adjustments/Cleaning method
<Cleaning of Detector>
Clean the pyrolyzer unit if it becomes extremely dirty. The pyrolyzer unit must be turned off while cleaning it.
Use a waste cloth to remove dust. Do not use water or organic solvent for cleaning because they may
cause malfunctions.
Because an extremely large amount of dust inside the tube may disturb the gas detection, it must be
cleaned with dry AIR, etc.
5-3. How to replace parts
<Replacement of Regular Replacement Parts>
List of recommended regular replacement parts
No. Item Maintenance
intervals
Replacement
intervals
Quantity
(pieces/unit)
1
Flow sensor
1 year
5 year
2
2
Fan
0.5 years
2 - 4 years
1
3 Pyrolyzer
heater
-
2 - 4 years
1
NOTE
•
The above replacement intervals are recommendation only. The intervals may change depending on the
operating conditions. These intervals do not mean the warranty periods either. The result of the regular
maintenance may determine when to replace the parts.
•
Replace the elbow at the same time as replacing the pyrolyzer heater.
Replacement of Flow Sensor, Fan, and Pyrolyzer Heater
After the flow sensor, fan, or pyrolyzer heater is replaced, the operation must be checked by a qualified
service engineer.
For the stable operation of the pyrolyzer unit and safety, ask a qualified service engineer to take care of
replacement of the parts that operation must be checked. Please contact our sales department.
5-4. Procedures to store the pyrolyzer unit or leave it
for a long time
The pyrolyzer unit must be stored under the following environmental conditions.
•
In a dark place under the normal temperature and humidity away from direct sunlight
•
In a place where gases, solvents or vapors are not present
CAUTION
When you use a relocated or stopped/stored detector again, do not forget to perform a gas calibration.
For information on readjustment including gas calibration, please contact our sales department.