9 Troubleshooting
9 Troubleshooting
- 48 -
<Abnormalities of Readings>
Symptoms Causes
Actions
The reading rises
(drops) and it remains
so.
Drifting of sensor
output
Perform the zero adjustment.
Presence of
interference gas
Disturbances by interference gases, such as solvents,
cannot be eliminated completely.
For information on actions, such as removal filter, please
contact our overseas sales department or local
representatives.
Slow leak
A very small amount of the gas to be detected may be
leaking (slow leak). Because ignoring it may cause
dangers, take a remedial measure, i.e., taking actions the
same as those for the gas alarm.
Environmental
changes
Perform the zero adjustment (AIR adjustment).
A gas alarm is triggered
despite of no gas leak
and no other
abnormalities at the
detection point.
Presence of
interference gas
Disturbances by interference gases, such as solvents,
cannot be eliminated completely. For information on
actions, such as removal filter, please contact our
overseas sales department or local representatives.
Disturbance by
noise
Turn off and restart the detector.
If such a symptom is observed frequently, take
appropriate measures to eliminate the noise.
Sudden change in
the environment
When the environment (temperature, etc.) changes
suddenly, the detector cannot adjust to it and is affected
by it.
In some cases, the detector triggers an indication alarm.
Because the detector cannot be used under sudden and
frequent environmental changes, any preventive actions
should be taken to eliminate them by the user.
Slow response
Clogged dust filter
Replace the dust filter.
Bended, clogged, or
leaked suction tube
or exhaust tube
Fix the defective parts.
Condensation is
formed inside the
suction tube.
Fix the defective parts.
Deteriorated sensor
sensitivity
Replace the sensor with a new one.
Sensitivity calibration
impossible
Improper calibration
gas concentration
Use the proper calibration gas.
Deteriorated sensor
sensitivity
Replace the sensor with a new one.
<Abnormalities of installation environment>
Symptom/Display FAULT
Causes
Actions
Out of operating
temperature
range
E-27
―
The environmental
temperature is out of
operating temperature
range
Operating temperature range
-20 to +60
℃
If environmental temperature is out of operating
temperature range, countermeasure for
adjusting environmental temperature is needed
to be considered.
Содержание GP MED EX
Страница 1: ...PT2E 3030 Smart transmitter Gas Detector Head SD 1D TYPE GP MED EX Operating Manual PT2E 161 ...
Страница 5: ......
Страница 6: ......
Страница 7: ......
Страница 8: ......
Страница 9: ......
Страница 10: ......
Страница 11: ......
Страница 13: ......
Страница 14: ......
Страница 16: ......
Страница 17: ......
Страница 18: ......
Страница 19: ......
Страница 20: ......
Страница 21: ......
Страница 22: ......
Страница 23: ......
Страница 24: ......
Страница 25: ......
Страница 26: ......
Страница 27: ......
Страница 28: ......
Страница 31: ......
Страница 32: ......
Страница 35: ......
Страница 36: ......
Страница 37: ......
Страница 38: ......
Страница 39: ......
Страница 40: ......
Страница 41: ......
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 46: ......
Страница 48: ......
Страница 52: ......
Страница 53: ......
Страница 56: ...Revision History Edition Revision Date issued 0 First edition 1 6 2021 ...
Страница 57: ......