Chapter 6 System Function
RIGOL
RSA5000 User's Guide
6-1
Chapter 6 System Function
System
Sets the system parameters.
Preset
Functions include: select the instrument settings to be recalled after the analyzer is
powered on ("Last" or "Preset"); set the preset type ("Factory" or one of the items
from "User1" to "User6"); save preset.
1. Pow er On
Sets the power on setting to "Last" or "Preset".
When "Last" is selected, the last system settings before power-off would be
recalled automatically at power-on.
When "Preset" is selected, settings def ined in
Preset Type
would be
recalled automatically at power-on.
2. Preset Type
Sets the preset type to "Default" (by default) or any one of the items from
"User1" to "User6".
When Power On is set to "Preset", the specif ied preset type will be recalled
at power-on.
After the instrument starts, press
Preset
on the front panel in any
operation interface will recall the specif ied preset type.
3. Save User Preset
Saves the current system setting as user-def ined settings to the internal
non-volatile memory. You can store up to 6 system states (correspond to
"User1" to "User6" in the preset type) and name each state f ile.
When one of the items from "User1" to "User6" is selected under
Preset Type
,
press
Save User Preset
, and the then automatically enter the f ilename input
interface. Then, perform the saving operation.