Chapter 2 Front Panel Operation
RIGOL
User’s Guide for DSA800/E Series
2-71
Table 2-45 Average Number for C/N Ratio Measurement
Parameter
Description
Default
10
Range
1 to 1000
Unit
N/A
Knob Step
1
Direction Key Step
1
2.
Avg Mode
Set the average operation mode to “Exp” or “Repeat” and the default is “Exp”.
When “Exp” is selected, the result is the exponential average of the
current N (N is specified in “Avg Num”) measurement results.
When “Repeat” is selected, the result is the arithmetic average of the
current N (N is specified in “Avg Num”) measurement results.
3.
Offset Freq
Set the difference between the carrier center frequency and the noise center
frequency. You can use the numeric keys, knob and direction keys to modify
this parameter.
Table 2-46 Frequency Offset for C/N Ratio Measurement
Parameter
Description
Default
2 MHz
Range
33 Hz to 2.5 GHz
Unit
GHz, MHz, kHz, Hz
Knob Step
Offset Frequency/100, the
minimum is 1Hz
Direction Key Step
in 1-1.5-2-3-5-7.5 sequence
4.
Noise BW
Set the bandwidth of the noise to be measured. You can use the numeric keys,
knob or direction keys to modify this parameter.
Содержание DSA815
Страница 1: ...RIGOL User s Guide DSA800 E Series Spectrum Analyzer March 2017 RIGOL TECHNOLOGIES INC...
Страница 2: ......
Страница 14: ...RIGOL User s Guide for DSA800 E Series XII...
Страница 49: ...Chapter 1 Quick Start RIGOL User s Guide for DSA800 E Series 1 29 Figure 1 14 The Built in Help Interface...
Страница 52: ......
Страница 189: ...Chapter 3 Remote Control RIGOL User s Guide for DSA800 E Series 3 5 5 6...
Страница 198: ...RIGOL Chapter 3 Remote Control User s Guide for DSA800 E Series 3 14 Figure 3 9 Write and Read Commands via GPIB...
Страница 238: ......