RIGOL
Chapter 2 Front Panel Operation
User’s Guide for DSA800/E Series
2-64
OBW
Measurement Interface:
Figure 2-17 OBW Measurement Interface
Measurement Results: occupied bandwidth and transmit frequency error.
Occupied Bandwidth: integrate the power within the whole span and then
calculate the bandwidth occupied by the power according to the specified
power ratio.
Transmit Frequency Error: difference between the center frequency of the
channel and the center frequency of the analyzer.
Measurement Parameters: average number, average mode, max hold, span
and power ratio.
1.
Avg Num
Specify the number of averages used when calculating the measurement
result and the default is “Off”. You can use the numeric keys, knob or direction
keys to modify this parameter.
Содержание DSA815
Страница 1: ...RIGOL User s Guide DSA800 E Series Spectrum Analyzer March 2017 RIGOL TECHNOLOGIES INC...
Страница 2: ......
Страница 14: ...RIGOL User s Guide for DSA800 E Series XII...
Страница 49: ...Chapter 1 Quick Start RIGOL User s Guide for DSA800 E Series 1 29 Figure 1 14 The Built in Help Interface...
Страница 52: ......
Страница 189: ...Chapter 3 Remote Control RIGOL User s Guide for DSA800 E Series 3 5 5 6...
Страница 198: ...RIGOL Chapter 3 Remote Control User s Guide for DSA800 E Series 3 14 Figure 3 9 Write and Read Commands via GPIB...
Страница 238: ......