5.4.3
Test Procedures
WARNING
Before connecting, disconnecting, or moving the test hookup, disable the output of the
signal generator to avoid causing the dangerous voltage.
1. The connection diagram is shown in the above figure:
a. Connect the [RF OUTPUT 50 Ω] connector of DSG3000B to the input of the
power splitter by using the N(M)-N(M) cable.
b. Connect the power meter probe (power sensor) to the output of the power
splitter.
c. Connect the other splitter output to CH1 of the oscilloscope by using the N(M)-
BNC(M) cable.
2. Set up the power meter: set the power meter to display measurements in watts.
3. Configure the oscilloscope:
a. Press the front-panel key
to enable CH1.
b. In the Vertical system menu, click or tap 50 Ω under Impedance to set the
input impedance of CH1 to 50 Ω. Set the vertical scale to 100 mV/div.
c. In the Horizontal system menu, set the horizontal timebase to 500 ns/div.
4. Output a Sine with 1 MHz frequency and 6 div amplitude (e.g. 5 mV/div vertical
scale, 30 mVpp Sine signal). Adjust the timebase to display five cycles of the
waveforms on the screen.
5. In the Measure menu, click or tap Vertical measurement item to select "AC.RMS".
The AC.RMS measurement result list is displayed at the right section of the screen.
Read the value from the "result" list and record the measurement result as Vout
1
MHz
.
6. Record the power meter reading P
1 MHz
, and convert the power to V according to
the following formula.
Performance Verification Test
DS70000 Performance Verification
24
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.