c. Click or tap the drop-down button of the BW Limit menu to select "20 M".
d. Set the vertical scale to 1 mV/div.
e. Set the offset to 1 V, as shown in the test record form. Close the Vertical system
menu.
f. Click or tap the horizontal menu ("D" icon) at the top of the screen. Then the
Horizontal system menu is displayed. In this menu, set the timebase to 1 ms/
div. Select "Average" for the Acquisition menu. Then click or tap the input field
for the Averages menu item to set it by using the pop-up numeric keypad. Set
the number of averages item to 16. Close the Horizontal system menu.
g. Click or tap the trigger information label at the top of the screen. Set Source to
AC Line.
h. Adjust the trigger level to avoid that the signals are being triggered by mistake.
4. Set the output of Fluke 9500B to -1 V.
5. In the Measure menu, click or tap Vertical measurement item to select "Vavg". The
Vavg measurement result list is displayed at the right section of the screen. Read
the value from the "result" list and record the measurement result.
6. Set the vertical offset of the oscilloscope to 0 V.
7. Set the output of Fluke 9500B to 0 V. Read the value from the "result" list and
record the measurement result.
8. Repeat Step 7-9. Measure and record the results according to
5.3.4
Test Record Form
Channel
Setting
Vertical
Scale
Offset
Test Result
Min.
Max.
CH1, 1 MΩ,
20 MHz BW
1 mV/div
1 V
-1.0171 V
-982.90 mV
1 mV/div
0 V
-2.1000 mV 2.1000 mV
1 mV/div
-1 V
982.90 mV
1.0171 V
200 mV/div 30 V
-30.472 V
-29.528 V
200 mV/div 0 V
-22.000 mV 22.000 mV
200 mV/div -30 V
29.528 V
30.472 V
500 mV/div 100 V
-101.05 V
-98.948 V
500 mV/div 0 V
-52.000 mV 52.000 mV
Performance Verification Test
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DS70000 Performance Verification
21