Figure 4.1 Self-calibration Menu
- Click or tap Start, and then the oscilloscope will start to execute the self-
calibration program.
- After starting the self-calibration program, click you can or tap Exit to cancel
self-calibration operation at any time.
3. After completing the self-calibration, restart the oscilloscope. In the Horizontal
system menu, select "Average" for the Acquisition menu. Then click or tap the
input field for the Averages menu item to set it by using the pop-up numeric
keypad. Set the number of averages to 16.
4. Set the vertical scale of each channel to 2 mV/div and view the offset of the
waveform of each channel. If the offset is greater than 0.5 div, check whether there
are interference signals around you and whether the power source is well
grounded. If yes, perform self-calibration again.
5. Click or tap Close to close the self-calibration information window.
4.2
Test Result Record
Record and keep the test result of each test. In the last chapter of this manual, a test
result record form is provided. The form lists all the test items and their
corresponding performance limits as well as spaces for users to record the test
results.
TIP
It is recommended that users photocopy the test record form before each test and record the
test results in the copy so that the form can be used repeatedly.
Overview
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DS70000 Performance Verification
9