![Rigol DP2000 Series Скачать руководство пользователя страница 95](http://html1.mh-extra.com/html/rigol/dp2000-series/dp2000-series_programming-manual_1461089095.webp)
Queries the current operation mode.
Parameter
Name
Type
Range
Default
<mode>
Discrete
{LOCal|REMote|RWLock}
LOCal
Remarks
•
LOCal: Local mode.
•
REMote: Remote mode. The touch screen and all front-panel keys are disabled
except for the front-panel output on/off keys and
key. At this point, you can
press the output on/off key only to disable channel output.
•
RWLock: Remote lock mode. The touch screen and all front-panel keys are
disabled except for the front-panel output on/off keys. At this point, you can
press the output on/off key only to disable channel output. You can only use the
specified command to disable the remote lock mode.
Return Format
The query returns LOCal, REMote, or RWLock.
Examples
:SYSTem:COMMunicate:RLSTate REMote /*Sets the power supply to
remote mode.*/
:SYSTem:COMMunicate:RLSTate? /*Queries the operation mode. The
query returns REMote.*/
4.11.7
:SYSTem:COMMunicate:RS232
4.11.7.1
:SYSTem:COMMunicate:RS232:BAUD
Syntax
:SYSTem:COMMunicate:RS232:BAUD
<
baud>
:SYSTem:COMMunicate:RS232:BAUD?
Description
Sets or queries the baud rate of the RS232 interface. The unit is Baud.