Parameter
Name
Type
Range
Default
<value>
Real
0.001 s to 3600 s
Refer to
Remarks
•
<value> specifies the duration of a cycle for a waveform. All waveforms except
for ExpRise and ExpFall require a period.
•
By default, the period of Pulse is 2 s, and the period of other waveforms is 50 s.
•
The total number of points within a period of the selected template is
determined by the period of the waveform and time interval
). The number of points within a period=period/time
interval.
Return Format
The query returns a real number ranging from 0.001 to 3600, for example, 15.000.
Examples
:TIME:TEMP:PERI 15 /*Sets the Period to 15 s.*/
:TIME:TEMP:PERI? /*Queries the Period for the selected template.
The query returns 15.000.*/
4.12.18 :TIMEr:TEMPlet:POINTs
Syntax
:TIMEr:TEMPlet:POINTs
<
value>
:TIMEr:TEMPlet:POINTs?
Description
Sets or queries the total number of points.
Parameter
Name
Type
Range
Default
<value>
Integer
1 to 512
50
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
113