Remarks
•
When OVP is enabled, the output is disabled automatically if the output voltage
reaches the OVP level currently set. You can send
or
to query whether an OVP event occurred on the
specified channel.
•
If <source> is omitted, the command sets or queries the OVP level of the current
channel.
•
You can select “MINimum” to set the minimum OVP level or “MAXimum” to
set the maximum OVP level.
•
[:SOURce[<n>]]:VOLTage:PROTection[:LEVel]
to set the OVP
level for the specified channel.
Return Format
The query returns the OVP level, for example, 8.800.
Examples
:OUTP:OVP:VAL CH1,8.8 /*Sets the OVP level of CH1 to 8.8 V.*/
:OUTP:OVP:VAL? CH1 /*Queries the OVP level of CH1. The query
returns 8.800.*/
4.8.14
:OUTPut:PAIR
Syntax
:OUTPut:PAIR
<
type>
:OUTPut:PAIR?
Description
Sets or queries the channel connection mode.
Parameter
Name
Type
Range
Default
<type>
Discrete
{OFF|PARallel|SERies}
OFF
Remarks
•
OFF: CH1 and CH2 are independent of each other.
•
PARallel: CH1 and CH2 are connected in parallel (internal).
•
SERies: CH1 and CH2 are connected in series (internal).
Return Format
The query returns the connection mode, for example, SERIES.
Command System
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.