Chapter 2 Front Panel Operations
RIGOL
DG800 User's Guide
2-25
frequency, an integer.
The maximum is 8.
You can also use the arrow keys and the knob to set the parameter value: press
down the knob to enter the editing mode, use the arrow keys to move the cursor to
select the digit to be edited, and then rotate the knob to modify the value. Besides,
you can use the knob to switch between different parameters.
To Select Harmonic Amplitude
After entering the harmonic setting interface, tap the Harm Ampl parameter input
field to set the amplitude of each order of harmonic.
1)
No.: selects the sequence number of the harmonic to be set.
2)
Harm Ampl: sets the amplitude of the specified order of harmonic. Use the
numeric keypad to input the amplitude value, and then select the desired unit.
Tap "Ok". The amplitude units available are Vpp, mVpp, Vrms, mVrms, and dBm
(invalid in HighZ).
To Set Harmonic Phase
After entering the harmonic setting interface, tap the Harm Phase parameter input
field to set the phase of each order of harmonic.
1.
No.: selects the sequence number of the harmonic to be set.
2.
Harm Phase: sets the phase of the specified order of harmonic. Use the
numeric keypad to input the phase value, and then select the desired unit "°".
Tap "Ok".
Example: To Output Harmonic
This section introduces how to output 2th order of harmonic (harmonic amplitude: 2
Vpp, harmonic phase: 30°) and 4th orders of harmonic (harmonic amplitude: 1 Vpp,
harmonic phase: 50°) from the [CH1] connector.
1.
Select the output channel: Press Output1 on the front panel or tap the
channel output configuration status bar
to select CH1. At this time,
the channel is indicated in red in the status bar.
Содержание DG800 Series
Страница 2: ......
Страница 14: ......
Страница 30: ......
Страница 46: ...RIGOL Chapter 2 Front Panel Operations 2 16 DG800 User s Guide Figure 2 3 Sine Waveform...
Страница 57: ...Chapter 2 Front Panel Operations RIGOL DG800 User s Guide 2 27 Figure 2 5 Harmonic...
Страница 136: ......
Страница 142: ......
Страница 148: ...RIGOL Index 2 DG800 User s Guide VISA Address 2 94 Waveform Combination 2 89 Width Deviation 2 55...