RIGOL
DG1000Z Performance Verification Guide
3
Harmonic Distortion Test
Frequency
Setting
Value
Setting
Measurement
Value
Calculation
Result
Specification Pass/Fail
10MHz
Waveform:
Sine
Amplitude:
0dBm
Offset:
0V
DC
Base waveform:
<-65dBc
2
nd
order
harmonic:
25MHz
Base waveform:
<-55dBc
2
nd
order
harmonic:
30MHz
[1]
Base waveform:
<-55dBc
2
nd
order
harmonic:
60MHz
[2]
Base waveform:
<-50dBc
2
nd
order
harmonic:
Note
[1]
: Only applicable to DG1032Z and DG1062Z.
Note
[2]
: Only applicable to DG1062Z.