Chapter 2 Performance Verification Test
RIGOL
DG1000Z Performance Verification Guide
2-19
4.
Turn on and set the oscilloscope:
a)
Set the vertical scale to 200mV/div.
b)
Set the horizontal time base to 1ns.
c)
Adjust the trigger level to a proper value.
d)
Set the input impedance to 50Ω.
e)
Turn on the rise time and fall time measurement functions.
5.
Set the edge type of the oscilloscope to rising edge, record the measurement
result of the rise time and judge whether it is within the specification range.
6.
Set the edge type of the oscilloscope to falling edge, record the measurement
result of the fall time and judge whether it is within the specification range.
7.
Repeat steps 1 to 6 to test the rise/fall time of CH2 and record the measurement
results.
Test Record Form:
Waveform Setting
Measurement Value
Specification
Pass/Fail
Square
Frequency: 1MHz
Amplitude: 1Vpp
Offset: 0V
DC
Rise Time
Typical (1Vpp)
<10ns
Fall Time