Chapter 2 Performance Verification Test
RIGOL
DG1000Z Performance Verification Guide
2-17
14.
Keep the reference level, input attenuation, resolution bandwidth, peak offset
and sweep mode of the spectrum analyzer as 10dBm, 20dB, 1kHz, 3dB and
single respectively. Set its start frequency to 0MHz and stop frequency to
150MHz.
15.
Repeat steps 8 and 9.
16.
Keep the output impedance of CH1 of DG1000Z at 50Ω. Set the output
waveform of CH1 as a sine waveform with 60MHz frequency, 0dBm amplitude
and 0V
DC
offset.
17.
Keep the reference level, input attenuation, resolution bandwidth, peak offset
and sweep mode of the spectrum analyzer as 10dBm, 20dB, 1kHz, 3dB and
single respectively. Set its start frequency to 0MHz and stop frequency to
300MHz.
18.
Repeat steps 8 and 9.
19.
Repeat steps 1 to 18 to test the spurious signal (non-harmonic) of CH2 and
record the test results.
Test Record Form:
Output
Frequency
Start
Frequency
Stop
Frequency
A
A-0dBm Specification Pass/Fail
5MHz
0Hz
30MHz
<-70dBc
10MHz
0Hz
50MHz
<-70dBc
20MHz
0Hz
100MHz
<-64dBc
25MHz
0Hz
150MHz
<-58dBc
30MHz
0Hz
150MHz
<-58dBc
60MHz
0Hz
300MHz
<-40dBc