Caution
Caution
- Before using the
Blood Glucose Monitoring System to test your blood glucose, please
RIGHTEST
read all of the instructions and conduct all of the tests including the quality control test.
- Please perform the quality control test regularly to make sure the test results are accurate.
-
Blood Glucose Meter can only be used with
Blood Glucose
RIGHTEST GT200
RIGHTEST GT200
Test Strips. Other brands’ test strips should not be used under any circumstances. The use of
other brands’ strips may give inaccurate results.
-
Blood Glucose Monitoring System is intended for in vitro diagnostic use only.
RIGHTEST
RIGHTEST Blood Glucose test results using fresh capillary whole blood samples from the
fingertip, palm and forearm are calibrated to be the equivalent to that of plasma testing.
-
Blood Glucose Monitoring System is intended for self-testing. It should not be used to
RIGHTEST
diagnose diabetes mellitus.
- Wait to test at least 30 minutes after entering another location of a different ambient temperature.
- Follow all environmental protection regulations when disposing of batteries.
- Keep the test strips, control solution or vial cap away from children. They may cause a choking
hazard. If a test strip or vial cap is swallowed, contact your physician immediately.
- Prevent water from entering the meter. Never immerse the meter or hold it under running water.
- A drop of blood is required to perform a blood glucose test. Capillary blood can be used, while
venous, neonatal or arterial blood may also be used only if drawn by healthcare professionals.
- The system has been tested with neonatal blood. As a matter of good clinical practice, be caution
of the interpretation of neonate glucose value below 50 mg/dL (2.8 mmol/L). Please follow the
recommendations for follow-up care that have been set by your intuition for critical glucose values
in neonates. Once patient is suspected as rare disease (galactosemia), the glucose result should
based on laboratory test.
- The minimum blood sample size to test using RIGHTEST Blood Glucose Monitoring System is
0.75
µL
: ( )
Blood sample size above 3.0
µL
might contaminate the test strip port and the meter.
Blood sample size below 0.75
µL
may cause an inaccurate result or may prevent a meter reading. An
Er4 will be displayed if the sample size is too small. In this case, repeat the test with a new test strip.
Sample Size Example
0.75
µL
1.0
µL
1.5
µL
2.0
µL
3.0
µL