Understanding Control Test Results
Example of control
solution range printed on
your test strip vial label
Control Solution Range
4.89 - 6.67 mmol/L
13.89 - 42.5 mmol/L
L2 = Level 2 (Normal)
L4 = Level 4 (High)
88 - 120 mg/dL
250 - 340 mg/dL
Recalling Test Results
Your control solution test results should fall within the control solution range. If the results are within
the range, RIGHTEST Blood Glucose Monitoring System is working correctly.
Possible reasons your Control Solution results are out of the range :
- Your RIGHTEST Control Solution is expired or was first opened more than 3 months ago.
- Your RIGHTEST Blood Glucose Test Strip has expired.
- You left the cap of the Blood Glucose Test Strip vial or the control solution off for a period of time.
- You did not perform the test procedure correctly.
- The RIGHTEST Blood Glucose Meter or RIGHTEST Blood Glucose Test Strip have malfunctioned.
If RIGHTEST Control Solution results are out of range, your RIGHTEST Blood Glucose Monitoring
System may not be working properly. Repeat the Quality Control Test. If your control solution results
are still out of range, do not use RIGHTEST Blood Glucose Meter to test your blood glucose. Please
contact customer service.
RIGHTEST Blood Glucose Meter is able to automatically store a maximum of 1,000 test results with
time and date. If your meter has stored 1,000 results, the newest test result will replace the oldest
one.
To recall your test memory, start the meter without a test strip inserted.
1. Press the Main button to switch from the Time Mode to Memory Mode. You
will see the "mem" icon in the upper left corner of the display. The latest test
result will appear with the number "1".
2. Use either the Left or Right button on the side to review all previous results
with date and time. Results display from the most recent (Sequence no. "1")
to the oldest (Sequence no. "1000") in the lower right hand corner of the
screen.
3. The Quality Control Test result can be recalled from the saved data. When
" " appears next to the data, the test was performed using the RIGHTEST
Control Solution. The control solution test result will not be used when
recalling test average results.