CAUTION
- Users should wash their hands thoroughly with soap and water after handling the meter,
lancing device or test strips.
- Please examine your LCD screen, test strip port, buttons and surface of your meter and
lancing device after cleaning and disinfecting cycles. Stop using the meter and/or
lancing device if any of the following occur:
- Thin, sliver streaks appear on the screen,
- The screen becomes cracked, soft, dissolved, brittle or swollen.
- You are unable to turn on/off your meter, operate the up/down button, the lancing
device release button or depth adjustable cap.
- You are unable to enter meter settings, function modes or recall your testing results.
If you have any questions or concerns, please contact your RIGHTEST Blood Glucose
Monitoring System authorized representative or contact your local Bionime distributor or
email to [email protected] .
Error Messages and Troubleshooting
Temperature Error
1. In order to get accurate testing results, please test between 6 - 44°C (43-111°F).
2. When the temperature is below 6°C (43°F) or over 44°C (111°F), the meter will
not function and the " " symbol will blink.
3. If the RIGHTEST Blood Glucose Meter and Test Strips are exposed to a
substantial change in temperature, please wait 30 minutes before measurement.
Maintain the Products