Performing a Test in Order
35
34
PRECAUTION
-
Load the lancing device ( Refer to Instruction manual of the lancing device )
- Be careful to load the lancing device to avoid being pierced by the needle.
Preparing your meter and strip
1. Remove new test strip from vial and
re-cap the vial cap immediately.
2. Insert the test strip with indication symbol, facing up, into test strip
port.
3. Push the test strip in until it snaps and stops. The meter turns on
automatically. When you put the strip into the strip port, you will hear
a sound of " beep ".
4. Check that the code number blinking on your meter matches the
code highlighted on the vial of test strips currently in use.
Performing a Test in Order
4
5
6
7
4) Twist off and set aside the protective cover of the disposable lancet.
5) Replace the depth adjustable cap.
6) Choose a depth of penetration by rotating the top portion of the depth adjustable
cap until the setting depth matches the window. Settings are based on skin type
" " for soft or thin skin; " " for average skin; " " for thick or
calloused skin.
7) Hold the hub in one hand and pull on the plunger in the other hand. The device will
be cocked. Release the plunger, it will automatically move back to its original
position near the hub.
5. While the " " symbol flashing, you are ready to apply the blood
sample. The code number will appear on the right down corner of the
screen.