4.  INSTALLING THE ROTOR TARGET IN AN X-RAY TUBE 

4.1  Unpacking 

 

(1)  Check the packing for breakage and other abnormality. 

 

(2)  After unpacking, check all supplied items with a LIST OF SUPPLIED ITEMS (section 3). 

 

(3)  Check for no flaw on the surface of the Rotor Target.    Stripes on the surface of the Rotor Target 

are the traces of electronic beams and do not constitute any problem. 

4.2    Mounting the Rotor Target in an X-ray Tube 

 

 

 

 

 

 

(1)  Carry the Rotor Target by securely holding the body of it. 

Never carry the Rotor Target by holding its cable or target drum.    In addition, do not give any 
shock to the target drum. 

 

(2)  Wipe the part of the Rotor Target body to be inserted into the X-ray tube with a cloth or paper 

moistened with hexane or like solvent.    Use care so that the solvent does not flow into the body.   
Do not directly touch the said part by your hand. 

 

(3)  Wipe the O-ring to be attached to the flange of the Rotor Target with a clean cloth or paper.   

Also, thoroughly wipe off any dirt from the O-ring groove of the body and the part of the X-ray 
tube at which the O-ring is engaged. 

 

(4)  Firmly hold the body, and slowly insert into position while aligning it with the X-ray tube 

mounting port.    Use extreme care so that the circumferential surface of the body does not bump 
against anything. 

 

(5)  Make sure that the flange on the Rotor Target body is in proper contact with the X-ray tube, and 

then tentatively tighten the four mounting screws by degrees and in turn. 

 

(6)  Properly connect the two connectors of the body. 

 

(7)  If a plug is attached to the cooling-water hose fitting, pull it out pushing the ring on the fitting. 

 

(8)  Properly connect the cooling-water hose to the body, aligning the label on the hose with the label 

on the body (by strongly pushing the tip of the hose into the fitting). 

 

(9)  Insert the leakage sensor to the fitting on a side of the body. 

 

(10)  Turn ON the circuit breaker and the main power.    After it starts emitting exhaust, firmly tighten 

the four mounting screws by degrees and in turn. 

Note:    Before attaching or removing the Rotor Target, be sure to turn OFF the X-ray 

generator power, wait for at least 30 minutes after it stops to emit exhaust and 
until the TMP revolutions fully drop to a speed safe to handle. 

Содержание 4364F501

Страница 1: ...Cat No 4364F501 DIRECT DRIVE TYPE ROTOR TARGET FXRT INSTRUCTION MANUAL Manual No ME14003C01...

Страница 2: ...ual has to be purchased for each product If there are any missing or incorrectly collated pages in the delivered instruction manual please contact the sales representative who you purchased the instru...

Страница 3: ...ray Tube 4 5 MAINTENANCE 5 5 1 Cleaning the Surface of a Target 5 5 2 Inspecting and Servicing the Cooling Water Seal 5 5 3 Failures Insufficient Vacuum Rotational Noise or Heavier Vibrations 5 5 4 N...

Страница 4: ......

Страница 5: ...be mounted in the high intensity X ray generators For the X ray generators in which the Rotor Target can be mounted and the other related equipment refer to their respective manuals It is also well t...

Страница 6: ...motor 5 Kind of target Target material Catalog number Cu Cr Mo 4364F501 4364B501 4364G501 Table 1 6 Maximum permissible load Target material Maximum permissible voltage kV Focus size mm2 Maximum permi...

Страница 7: ...or Target 1 The body is supplied with the water leakage prevention plug installed over the cooling hose fitting 2 Spare parts The following spare parts are contained with the body in a storage box Car...

Страница 8: ...the body and the part of the X ray tube at which the O ring is engaged 4 Firmly hold the body and slowly insert into position while aligning it with the X ray tube mounting port Use extreme care so th...

Страница 9: ...commended to replace the carbon ring at intervals of 3000 hours The warranty is good for a period of 3000 hours or one year after delivery whichever occurs first 5 3 Failures Insufficient Vacuum Rotat...

Страница 10: ...2 In replacement of a target use care so that water drops remaining within the cooling water hoses do not come into contact with the target surface or the inside of the body of the Rotor Target 3 It i...

Страница 11: ...ining screws whether properly tightened There is an alarm issued for the target driver Solve the cause of the alarm then turn the power OFF on the operation panel and turn it ON again The target does...

Страница 12: ...get needs to be repaired or serviced in our factory to secure the performance it possesses at shipment Thus if the Rotor Target is in need of repair or service contact our service department See the f...

Страница 13: ...s supplied 8 A347 3 2D SiC ring 1 30 9 5661 48 3D Carbon ring 1 30 One piece supplied 10 5661 4865 Sheet cushion 1 30 t1 11 C048 304 Seal case 1 61 12 C048 3 12 O ring 1 S15 Viton 13 4555 307 Key 1 L5...

Страница 14: ......

Страница 15: ...tion Rigaku Americas Corporation Rigaku Europe SE Head Office 3 9 12 Matsubara cho Akishima shi Tokyo 196 8666 Japan Phone 81 42 545 8189 Fax 81 42 544 9223 e mail rinttyo rigaku co jp 9009 New Trails...

Отзывы: