Page 78 of 79
Technical Documentation and
User Instructions
Laser Mirror Scanner LMS-Q280i
Error
Err
No
Meaning
mFATAL:BAD_STREAM_ID
54
Internal Error in downloading hardware
specific data
mFATAL:SSL_TEMP_OUT_RANGE
57
Solid state laser types only: temperature
out of range
mFATAL:SSL_SHUTTER FAILED
58
Solid state laser types only: shutter function
failed
mERROR:SCAN_COMMUNICATION
29
Error in communication between range
finder unit and scanner unit. See parameter
SC_ERROR for detailed reason of error
mERROR:SCAN_STATUS
30
Error in scanner unit. See parameter
SC_STATUS for detailed reason of error
mERROR:SCAN_NO_MOTION
28
Error, frame scanner does not move.
mFATAL:SID_RW_FAILED
70
Internal Error in Serial Interface Device
mERROR:PRR_OUT_OF_RANGE
71
Solid state laser types only: Laser Pulse
rate out of range
mFATAL: PELTIER_TEMP_DEFECT
72
Solid state laser types only: Internal
hardware fault (Peltier element)
mFATAL: PELTIER_ DEFECT
73
Solid state laser types only: Internal
hardware fault (Peltier element)
mERROR: PRR_INIT_OUT_OF_RANGE
74
Solid state laser types only: Internal
hardware fault: Laser Pulse rate out of
range
mFATAL:ECP_OVERFLOW
79
Error on internal ECP data port; ECP port
fifo overflows
mWRNG:ECP_PORT_OPEN
82
Several parameters (e.g. the number of
measurements per line RF_NUMBER_L)
are displayed in the ECP port Data Header.
When changing such parameters, the
parallel ECP port must be closed the
parameters changed and programming
mode left, and the data port newly
reopened, so the new changed header can
be read by the user application. Therefore,
close the ECP parallel port before changing
such parameters.
mFATAL:ECP_SELFCHECK_FAILED
97
Error on internal ECP and / or TCPIP
hardware: internal self check failed
mERROR:TCPIP_BAD_MAC_ADDR
90
Ethernet MAC address invalid
mERROR:TCPIP_BAD_IP_ADDR
91
IP address invalid
mERROR:TCPIP_NOT_CONFIGURED
92
Internal LAN interface configuration error
mWRNG:TCPIP_DATA_PORT_OPEN
93
Several parameters (e.g. the number of
measurements per line RF_NUMBER_L)
are displayed in the Data Header. When
changing such parameters, the connection
to TCP/IP data port must be closed, the
parameters changed and programming
mode left, and the data port newly
reopened, so the new changed header can
be read by the user application. Therefore,
close the TCP/IP data port before changing
such parameters.