4

7. 

Adjusting the aperture diaphragm and condenser

 

The aperture diaphragm decides the numerical aperture (N.A.) of the illumination system.  If the N.A.   

 

of the illumination system matches with the NA of the objective, it can obtain better resolution and    

 

contrast and increase the depth of field.

 

a. 

You can adjust the condenser adjustment knob by turning it towards you. Raise the condener

 

 

towards the slide on the stage to fill the field of view with more light or turn the knob away 

 

 

from you to lower the condenser and fill less of the field of view with light. Since natural light 

 

 

to your specimen is usually low, it is a good idea  to adjust the condenser aperture diaphragm

 

 

to be 70-80% of the NA of the objective.  Adjust the aperture adjustment lever to allow the 

 

 

correct amount of light through. A good rule of thumb is  when using the 4x objective move 

 

 

the lever all the way to the right (open), and when using the 100x objective move the lever all

 

 

the way to the left (closed). For the objectives in between adjust the lever betwen these 

  settings.

8. 

Using the oil immersion objective

 

a.  

First, use the 4x objective to focus the specimen.

 

b. 

Place a drop of oil on the specimen.  

 

c. 

Rotate the nosepiece counterclockwise and rotate the oil objective (100x) into the light path.    

 

 

Then use the fine focusing knob to focus the specimen. 

 

 

**Make sure there are no air bubbles in the oil.  Air bubbles will distort the image. If you do 

 

 

experience an air bubble, you can gently swing the oil objective a few times to remove the 

  bubble.

 

d. 

After use, wipe the lens with a tissue moistened with lens cleaner.  Make sure all excess oil is    

 

 

removed from the objective after use.  

Left over oil can damage the objective if not properly  

 

  removed.

9. 

Use of colored filters

 

Swing out the colored filter holder at the bottom of the condenser 

 

(1) in a clockwise direction.  Place the filter (2) into the hole and 

 

then replace the holder.

      

      10. 

Replacing 

the 

Fuse

 

 

 

 

 

 

Turn the main switch to OFF before replacing the fuse.  Pull out 

 

 

 

 

 

 

the power cord.  Pull the fuse base  until the fuse can be 

 

 

 

 

 

 

removed from the fuse base. Install new fuse and push it into the

      fuse 

base 

firmly.

Содержание UX1B

Страница 1: ...Richter Optica info richter optica com Instructions for Models UX1B Binocular and UX1T Trinocular Laboratory and University Biological Microscope...

Страница 2: ...cular Eyepiece Tube Stage Specimen Holder Objectives Condenser Focusing Knob Coarse Focus Knob Fine Focus Knob Illumination on off switch Light Intensity Control Stage Y Axis Travel Knob Stage X Axis...

Страница 3: ...fingerprints or oil Damage of lens surfaces occur when they come in contact with such contaminants and image quality is reduced BEFORE USE When moving the microscope use both hands and only lay on fl...

Страница 4: ...he silver plate on the base of the microscope To access replace the batteries remove all screws in the corners of the silver metal plate Once screws are removed lift off metal plate and access battery...

Страница 5: ...objective While looking through the eyepieces move the eyetubes either further apart or closer together until the right and left field of view become one Line up the dots 1 on either side of the basep...

Страница 6: ...o the left closed For the objectives in between adjust the lever betwen these settings 8 Using the oil immersion objective a First use the 4x objective to focus the specimen b Place a drop of oil on t...

Страница 7: ...agm closed too far No cover glass Too thick or thin cover glass Immersion oil not used on immersion procedure Air bubbles in immersion oil Immersion oil used on a dry objective Greasy residue on eye l...

Страница 8: ...eyepiece differ Inadequate illumination Lamp does not light Inadequate brightness Lamp blows out immediately Lamp flickers Power supply not plugged in Lamp not installed Lamp burnt out Specified lamp...

Страница 9: ...tage is fashioned incorrectly Make sure cover glass is facing up Use a standard cover glass with a thickness of 0 13 0 17mm Turn the nosepiece into the right position until you hear it click into plac...

Страница 10: ...t use any solvents other than absolute alcohol as they may damage lens bonding cement 3 Do not use petroleum benzene when cleaning components such as filters or lenses 4 Absolute alcohol and petroleum...

Отзывы: