Richter Optica UX1B Скачать руководство пользователя страница 5

3

GETTING STARTED 

 

 

 

 

2. Placing the Specimen Slide

 

 

 

 

Push the silver finger of the specimen holder away from you (1).  Place the 

 

 

 

 

specimen slide (2) against the other side of the specimen holder while holding the 

 

 

 

 

finger open.  Once slide is placed properly with the cover glass facing upward you can 

 

 

 

 

release  the silver finger so that the slide remains clamped. Use the x and y axis adjust

 

 

 

 

ment knobs (3) to manuever the slide into the proper position where it is centered 

    above 

the 

light 

source. 

 

3. 

Adjust Focusing 

 

 

a. 

Shift the 4x objective into the light path until you hear it click properly in place.

 

 

b. 

Observe with just the right eye in the right eyepiece.  Rotate the coarse focusing knob until the 

 

 

 

image appears clearly in the field of view.

 

 

c. 

Rotate the fine focus knob to achieve finer focused details.

 

2. 

Coarse Focus Tension Adjustment 

 

 

 

To increase the tension, turn the tension adjustment ring located on the inside of the coarse focus knob  

 

 

in the direction indicated by the arrow. 

 

3. 

Interpupillary Distance Adjustment 

 

 

 

Before adjusting the interpupillary distance, bring a specimen into 

 

 

focus using the 4x objective. While looking through the eyepieces 

 

 

move the eyetubes either further apart or closer together until  the 

 

 

right and left field of view become one. Line up the dots (1) on either 

 

 

side of the basepoint scale (2). This adjustment will enable the user 

 

 

to see the specimen with both eyes.

 

4. 

Diopter adjustment

  

Diopter adjustment will compensate for the differences in vision between the left and right eyes.  In  addition  

 

 

to making observation with both eyes easier, it will also help reduce the extent to which focusing is lost when  

 

 

the objective magnification is changed.  In particular, this occurs when a low magnification objective is used.

 

 

Rotate the nosepiece until the 4x objective “clicks” into position. Using the fine and/or coarse focus  

 

 

 

knobs, focus the microscope until the image is at its sharpest. The left and right eyepieces have 

 

 

separate focusing provisions to compensate for slight differences in the focusing of each eye.

 

 

Using the right eye only and viewing through the right-hand 

 

 

eyepiece, adjust the diopter until the image is clearly in focus.

  

Using the left eye only and viewing through the left-hand 

 

 

eyepiece with its independent diopter focusing ring, focus 

 

 

until the specimen is at its sharpest. Once completing these 

 

 

steps, the microscope should now be ready for binocular viewing.

Содержание UX1B

Страница 1: ...Richter Optica info richter optica com Instructions for Models UX1B Binocular and UX1T Trinocular Laboratory and University Biological Microscope...

Страница 2: ...cular Eyepiece Tube Stage Specimen Holder Objectives Condenser Focusing Knob Coarse Focus Knob Fine Focus Knob Illumination on off switch Light Intensity Control Stage Y Axis Travel Knob Stage X Axis...

Страница 3: ...fingerprints or oil Damage of lens surfaces occur when they come in contact with such contaminants and image quality is reduced BEFORE USE When moving the microscope use both hands and only lay on fl...

Страница 4: ...he silver plate on the base of the microscope To access replace the batteries remove all screws in the corners of the silver metal plate Once screws are removed lift off metal plate and access battery...

Страница 5: ...objective While looking through the eyepieces move the eyetubes either further apart or closer together until the right and left field of view become one Line up the dots 1 on either side of the basep...

Страница 6: ...o the left closed For the objectives in between adjust the lever betwen these settings 8 Using the oil immersion objective a First use the 4x objective to focus the specimen b Place a drop of oil on t...

Страница 7: ...agm closed too far No cover glass Too thick or thin cover glass Immersion oil not used on immersion procedure Air bubbles in immersion oil Immersion oil used on a dry objective Greasy residue on eye l...

Страница 8: ...eyepiece differ Inadequate illumination Lamp does not light Inadequate brightness Lamp blows out immediately Lamp flickers Power supply not plugged in Lamp not installed Lamp burnt out Specified lamp...

Страница 9: ...tage is fashioned incorrectly Make sure cover glass is facing up Use a standard cover glass with a thickness of 0 13 0 17mm Turn the nosepiece into the right position until you hear it click into plac...

Страница 10: ...t use any solvents other than absolute alcohol as they may damage lens bonding cement 3 Do not use petroleum benzene when cleaning components such as filters or lenses 4 Absolute alcohol and petroleum...

Отзывы: