Page 7

FOCUSING TENSION ADJUSTMENT:

Included with the microscope is a focus 

tension adjustment tool. If the stage starts to 

drift (slowly fall) and samples are falling out of 

focus, the tension adjusment needs to be 

adjusted. Insert the tool into the slot on the 

coarse focus knob and rotate.

Turning one direction will tighten the tension 

and turning the opposite will loosen the 

tension.

If it becomes hard to turn the focus knob, 

loosen the tension slightly with the tool.

Содержание MDS1

Страница 1: ...Richter Optica info richter optica com Instructions for Model MDS1 Compound Microscope...

Страница 2: ...MDS1 Compound Microscope 10x Eyepiece Carrying Handle Rotatable Viewing Head Objective Lens Stage Clip Stage Built in Condenser Lens LED Light Coarse Focus Fine Focus Rheostat Control Page 1...

Страница 3: ...MDS1 Compound Microscope Page 2 Power Cord Plugs in Here On Off Switch Carrying Handle Rheostat Control for Light Slot for Adjusting Focusing Tension...

Страница 4: ...ECE Lens closest to the eye Magnifies the primary image formed by the objective lens Microscope includes 10x FN 16mm eyepiece B OBJECTIVE LENS Lens closest to the specimen forms the first magnified im...

Страница 5: ...ion to another requiring little or no adjustment of the fine focus knobs e As you increase the magnification the field of view area of specimen seen through the microscope will decrease That is why it...

Страница 6: ...ube length 132mm Parfocal Distance 30mm Monocular inclined viewing head 45 Included Eyepiece WF10x 16mm Illumination LED Rechargeable MICROSCOPE OPERATION Place the microscope on a stable work surface...

Страница 7: ...nses Clean the microscope surface with a soft cloth and for tough stains use a neutral detergent Do NOT use solvents such as alcohol or ether as this may damage the paint on the body of the microscope...

Страница 8: ...y fall and samples are falling out of focus the tension adjusment needs to be adjusted Insert the tool into the slot on the coarse focus knob and rotate Turning one direction will tighten the tension...

Страница 9: ...rp Cover slip is not in place Apply cover slip to slide Cover slip is too thick Use standard 0 17mm cover slip Prepared slide is upside down Turn slide so cover slip faces up Dirt on lenses Clean eyep...

Отзывы: