Guidance and manufacturer’s declaration - electromagnetic emissions.

The device is intended for use in the electromagnetic enviroment specified below. The 

customer or the user of the device should assure that it is used in such an environment.

Immunity 
test

IEC 60501 
test level

Compliance 
level

Electromagnetic environment  guidance

Conduct-
ed
RF lEC
61000-
4-6

3 Vrms
150 kHz
to 80 MHz

3 Vrms

Portable and mobile RF Communications 
equipment should be used no closer to any 
part of the device, including cables, than 
the recommended separation distance 
calculated from the equation applicable 
to the frequency of the transmitter. 
Recommended separation distance:

Where P is the maximum output power 
rating of the transmitter In watts (W) 
according to the transmitter manufacturer 
and d is the recommended separation 
distance in meters (m). Field strengths 
from fixed RF transmitters, as determined 
by an electromagnetic site survey, a should 
be less than the compliance level in each 
frequency range b. Interference may occur 
in the vicinity of equipment marked with the 
following symbol.

Radiated
RF lEC
61000-
4-3

3 V/m 80
MHz to
2.5 GHz

3 V/m

NOTE I. At 80 MHz and 800 MHz. the separation distance for the higher frequency range applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation 
is affected by absorption and reflection from structures, objects and people.

1. Field strengths from fixed transmitters, such as base stations for radio (cellular/

cordless) telephones and land mobile radios, amateur radio, AM and FM broadcast 

and TV broadcast cannot be predicted theoretically with accuracy. To assess the 

electromagnetic environment due to fixed RF transmitters, an electromagnetic site 

survey should be considered. If the measured field strength in the location in which 

the device is used exceeds the applicable RF compliance level above, should be 

observed to verify normal operation. If abnormal performance is observed, additional 

measures may be necessary, such as reorienting or relocating the device.

2. Over the frequency range 150 kHz to 80 MHz, field strength should be less than 

(3i) W/m.

80 MHz    
to 800  MHz
800 MHz   
to 2.5 MHz 

Содержание US 1000 3RD EDITION

Страница 1: ...INSTRUCTION MANUAL US 1000 3RD EDITION...

Страница 2: ...r notice Amendments may however be published in new editions of this manual All Right Reserved Rev V2 2019 United States Federal Law restricts this device to sale by or on the order of a physician or...

Страница 3: ...NS _______________________________________ 9 4 1 Accessories 4 2 Technical information 5 INSTRUCTIONS FOR USE ________________________________ 11 5 1 Function Test of Ultrasonic Action 6 MAINTENANCE A...

Страница 4: ...US 1000 is a Therapeutic ultrasound device that generates pulsed high frequency sound waves 1MHz that are transferred to a specific body are via a sound headed probe The Pulsed Sound Waves travel dee...

Страница 5: ...DO NOT use for patients with demand type cardiac pacemakers 7 DO NOT use on someone who is pregnant 8 DO NOT use over ischemic tissues in patients with vascular disease where the blood supply would b...

Страница 6: ...device away from children 6 DO NOT use over an area of the spinal cord following laminectomy i e when major covering tissues have been removed 7 Avoid bony prominences 8 When using ultrasound keep th...

Страница 7: ...tor or any other metallic or electronic device should not use this device without first consulting a doctor 9 Patients with the following diseases or symptoms should not use the device Pregnancy or Me...

Страница 8: ...r the patient with bleeding physique DO NOT use US 1000 WARNING The device complies completely with all parts of 21 CFR 1050 10 of the performance standard for sonic infrasonic and ultrasonic radiatio...

Страница 9: ...LED INTENSITY ON OFF MAIN BODY INDICATOR HI MED LO BUTTON TREATMENT HEAD MAIN BODY 4 SPECIFICATIONS 4 1 Accessories NO DESCRIPTION QTY 1 Ultrasound Device DU1025 1 2 Wall Adapter DU1025X 1 3 Therason...

Страница 10: ...Humidity Dimensions 6 80 x 2 25 x 1 65 inches L W H Weight 4 25 ounces Frequency 1MHz 10 Type Non focusing Pulse Width 8 0 5ms 10 14 1ms 10 29 2ms 10 Repetition Rate 150 Hz 10 Temporal Maximum Power 4...

Страница 11: ...s and will void warranty Apply Transmission Gel Wash the area to be treated so that it is free of oil and dirt Apply a generous layer of ultrasound transmission gel on the treatment area The gel acts...

Страница 12: ...er shut down press the On Off button and hold down for 4 seconds and the device will turn off 5 1 Function Test of Ultrasonic Action Place the probe horizontally then apply several water drops on the...

Страница 13: ...ected and tested by a Service Technician certified by an Authorized Service Center Cleaning the Applicator The applicator should be regularly inspected for damage e g hairline cracks which could allow...

Страница 14: ...cal distributor or the manufacturer as shown on the label PROBLEM CHECK POINTS POSSIBLE SOLUTION LED light does not turn on 1 The plug adapter is not inserted into the socket properly 2 The DC plug of...

Страница 15: ...interferences as well as maximum levels of electromagnetic emissions for medical devices Medical devices manufactured by Compass Health Brands Corp conform to this IEC50501 1 2 2007 standard for both...

Страница 16: ...up 1 The device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11...

Страница 17: ...000 4 4 2 kV for power supply 2 kV for power supply Main power quality should be that of a commercial or hospital environment If the user of the device requires continued operation during power mains...

Страница 18: ...frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF lEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m NOTE I At 80 MHz and 800 MHz the separati...

Страница 19: ...e according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 37 0 37 0 74 1 1 17 1 17 2 33 10 3 69 3 69 7 38 100 11 67 11 67 23 33 For tra...

Страница 20: ...sposed of with household waste after their useful life Help us to protect the environment and save resources and take this device to the ap propriate collection points Please contact the organization...

Страница 21: ...manufacturer to the consumer or during transport to the retailer Accessories which are subject to normal wear and tear Liability for direct or indirect consequential losses caused by the unit is exclu...

Страница 22: ...and return freight charges will apply All returns of dropshipped items are subject to a restocking fee as well as inbound and outbound freight charges 7 Returns will not be accepted on items that are...

Страница 23: ...23...

Страница 24: ...Manufactured for 3289 19 06 A 2019 Compass Health Brands Corp...

Отзывы: