Begin Treatment
Use gentle upward or circular motions with
the treatment head on treated area. Keep the
treatment head gliding over the skin,
CAUTION
DO NOT
stop moving the treatment head as it
could cause potential burns or injury.
Turning Off the Device
The device will automatically shut down after
working 30 minutes with 5 beeps. To turn off
power prior to the auto timer shut down, press
the On/Off button and hold down for 4 seconds
and the device will turn off.
5.1 Function Test of Ultrasonic Action
Place the probe horizontally, then apply several water
drops on the surface of the probe. Turn on the device,
you can observe the ultrasonic action. The water drops
on the probe start to perform one million vibrations
per second with slight atomization phenomenon.
Intensity
LOW: The water drops vibrate slightly.
MED: The water drops vibrate stronger, and with atomization.
HI: The water drops vibrate very strongly, and with atomization.
ON/OFF
L.M.H