Operation

This section describes the front panel and includes procedures for operation of the scale.

Figure 3-1. Front Panel Keypad

3.1 Key Descriptions

The display has 10 front panel keys. Key functions are described in the table below.

  The front panel keys are very sensitive, so only a gentle press is required.

Key

Name

Function

On/Off

Powers the scale on or off

Print 

LB/KG

Sends data out from the RS-232 port; Allows to toggle between kilograms and pounds providing that it is enabled in 

Configuration

 mode; Cannot toggle while in the 

BMI

 mode

Zero

Only functions if the current weight is stable and less than 2% of the capacity of the scale. Anything over 2% requires 

a recalibration

Hold Release Displays most current weight value on the display and holds that value when the patient is off the scale. A second 

press releases the weight value. Not active while in 

BMI

 mode

BMI

Pressing the 

BMI

 key enables access to the BMI (Body Mass Index) mode (defaults when scale is turned on). The 

patient is gets on the scale, weight stabilizes and press the 

BMI

 key. The display then asks for the patient height to 

calculate out the patient BMI.   

TARE

Used to remove the weight initially of anything on the scale that shouldn’t be included in the total weight of the 

patient on the scale 

CLEAR

When using the BMI function, the display looks for a height entry. Pressing 

Clear

 changes this entry back to 190.0 

cm (default) or 5 ft, 7.5 in.Once BMI is displayed, pressing the 

Clear

 key exits BMI

ENTER

Used to accept height in 

BMI

 mode; accepts the value of the parameter last entered and moves to the next stage

Pressing and holding Enter during startup will display ID. This is the first setup on entering into configuration mode

Up Arrows

Adjusts the value of the flashing digit/number

Adjusts height input (0.5 in/0.5 cm) while in 

BMI

 mode

Down Arrows Adjusts the value of the flashing digit/number

Adjusts height input (0.5 in/0.5 cm) while in 

BMI

 mode

Table 3-1.  Key Functions

IMPORTANT

On/Off

LB/KG

Print

Zero

Hold

Release

BMI

TARE

T

CLEAR

ENTER

Содержание Summit SD-1150-WP

Страница 1: ...Summit SD 1150 WP Dialysis Wheelchair Scale Software Version 11525 Operation Manual PN 197294 Rev A August 18 2022...

Страница 2: ...respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserves the right to...

Страница 3: ...ing or obtained by calling 715 234 9171 and asking for the training department Revision History This section tracks and describes the current and previous manual revisions for awareness of major updat...

Страница 4: ...4 Assembly 4 2 5 Electrical Interface to the Indicator 4 2 5 1 Insert Batteries 5 2 5 2 Power Connection 5 2 6 Load Cell Connections 6 2 7 Pit Installation 7 2 8 Dimensions 7 3 0 Operation 8 3 1 Key D...

Страница 5: ...nst harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance wit...

Страница 6: ...rt the scale while someone is on the scale Do not allow minors children or inexperienced persons to operate this scale Do not use in the presence of flammable materials Do not use this product if any...

Страница 7: ...the indicator must be protected from crushing cutting and moisture damage The scale must not be loaded beyond capacity even momentarily Choose a site where the floor is level to 1 4 inch to avoid exc...

Страница 8: ...cess plate located on the side of the scale Use the following steps to wire the junction box 1 Remove the two 10 x 3 8 inch screws 2 Slide the junction box assembly out of the deck 3 Open and remove t...

Страница 9: ...4 Battery Chamber 3 Put the cover in place and turn the thumbscrew clockwise to secure Remove the batteries prior to storing if the product is not going to be used for an extended period of time If t...

Страница 10: ...to gain access to the load cell connection Figure 2 6 Remove Tilt Stand 2 Remove the four back retaining screws to remove the back cover to the indicator Figure 2 7 Remove Back Cover 3 Plug the end o...

Страница 11: ...ale in a floor level pit and are for reference only the pit must be installed in poured concrete foundation according to standard construction practices Figure 2 9 Pit Frame Diagram A 1 24 slope is re...

Страница 12: ...MI Pressing the BMI key enables access to the BMI Body Mass Index mode defaults when scale is turned on The patient is gets on the scale weight stabilizes and press the BMI key The display then asks f...

Страница 13: ...y the patient 1 Press to turn on the scale 0 0 appears on the display along with the ZERO annunciator 2 Place additional item s on the scale 3 Press until the display returns to 0 0 and NET annunciato...

Страница 14: ...adjust the height value 5 Press 6 The BMI value and BMI annunciator are shown on the display Press to return to the Weighing mode and the BMI function will be turned off 3 6 2 KG Mode 1 Ensure that t...

Страница 15: ...high E07 AD too low E10 Overload Scale has been overloaded Remove load from scale E4L BAT Battery low but still usable one bar left on indicator display E4U Battery low and unstable no bars left on in...

Страница 16: ...ess and hold for at least three seconds or until the scale emits two quick beeps If the scale does not beep after five seconds release as the weight was either in motion or out of range If displaying...

Страница 17: ...the use of the USB driver Driver protections may need to be temporarily disabled on Windows 10 or later computers to allow for the installation of the USB driver 1 Connect the scale s indicator to a...

Страница 18: ...the overall appearance of the entire scale for any obvious signs of damage Inspect the condition of the AC power adapter cord for cracking fraying or for broken or bent prongs 5 2 Cleaning Proper car...

Страница 19: ...urs of continuous use Automatic power off can be configured Environmental Operating Temperature 50 F to 104 F 14 C to 40 C Storage Temperature 32 F to 158 F 0 C to 70 C Humidity 85 relative humidity D...

Страница 20: ...Summit SD 1150 WP Wheelchair Scale 16 Visit our website www RiceLake com...

Страница 21: ......

Страница 22: ...54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems Specifications subject to change without notice www ricelake com PN...