SCT Weight Transmitter – 20 Series Technical Manual
40
Visit our website
Status Register
Table 3-25. Status Register (EtherNet/IP)
Bit No.
Description
Bit 0
Cell Error
Bit 1
A/D Converter Malfunction
Bit 2
Maximum weight exceeded by 9 divisions
Bit 3
Gross weight higher than 110% of full scale
Bit 4
Gross weight beyond 999999 or less than -999999
Bit 5
Net weight beyond 999999 or less than -999999
Bit 6
Bit 7
Gross weight negative sign
Bit 8
Net weight negative sign
Bit 9
Peak weight negative sign
Bit 10
Net display mode
Bit 11
Weight stability
Bit 12
Weight within ±¼ of a division around ZERO
Bit 13
Bit 14
Bit 15
3.9
SCT20-TCP/IP (Ethernet TCP/IP)
To configure Ethernet TCP/IP port of SCT40-TCP/IP, the Lantronix DeviceInstaller must be installed on a Windows PC.
1. Connect the PC and SCT20-TCP/IP through a LAN, point-to-point or with hub/switch.
2. Run the Lantronix Device Installer file (DevInst.exe) on the CD.
3. Select
Search
and the application will find the SCT20-TCP/IP in the LAN.
Figure 3-2. Find SCT20-TCP/IP in LAN
4. Select the device and then select the
Telnet Configuration
.