12
RL-650 Series Operation Manual
6.0
Troubleshooting & Testing
Refer to the following instructions to check and correct any failure before consulting service personnel.
Symptom
Possible Cause
Corrective Action
The scale does not turn on.
Dead batteries
Connect the scale to a power source or
change batteries.
Faulty electrical outlet
Use a different electrical outlet.
Bad power supply
Replace optional adapter.
Questionable weight or the scale does
not zero.
External object is interfering with the
scale.
Remove the infant/interfering object from
the weighing tray from the scale.
Display did not show 0.0 before weighing. Remove the infant, zero the scale and
begin the weighing process again.
Scale is not placed on a level surface.
Place the scale on a stable surface and
begin the weighing process again.
The weighing tray is not placed properly.
Place the weighing tray in its proper
place.
Scale is out of calibration.
Check the scale with a known weight
value.
The display shows
o_Ld
message.
The load on the scale exceeds the
capacity.
Remove the excess weight and use the
scale according to its stated limit.
The display shows Err.
The
Recall
key was pressed with
insufficient weight on the scale.
Place more than 2 lb, 30 oz on the scale.
Table 6-1. Troubleshooting .