8

RL-650 Series Operation Manual

4.5

Scale Operation with Manual Tare

Use the following steps to use the manual tare function.

1. Press the 

On/Off

 key to turn the scale on.

2. When the weight is set to zero, place the item to tare (ie: blanket) on the scale. Press the 

Tare

 key until the

display returns to zero and the 

Net

 annunciator appears on the display.

3. Remove the item (ie: blanket) from the scale. The weight will appear as a negative weight.
4. Place the infant back on the scale with the previously tared item. The display shows the infant’s weight.

The 

Net

 annunciator is still on. Clear the scale of all items and the weight of the extra load remains stored

in memory.

5. To cancel the tare weight, press the 

Tare

 key until 

Net

 disappears from the display. The display turns back

to zero and the 

Gross

 appears. The tare weight is also canceled when the scale is turned off. 

4.6

 Preset Tare

Use the following steps to use the preset tare function.

1. Press the 

On/Off

 key to turn the scale on.

2. Wait until the zero appears on the display and press the 

Tare

 key. 

3. The default tare value is displayed (default is programmed to be 33.0 lb/15.0 kg) while the zero is 

flashing.

4. Use the 

Up/Down

 and 

left arrow

 keys to adjust the value.

5. Press the 

Enter

 key to start the tare function and the 

Net

 annunciator will be shown instead of the 

Gross

 

annunciator. 

Содержание RL-650 Series

Страница 1: ...PN 163975 Rev B RL 650 Neonatal Scale 650 10 1 Operation Manual...

Страница 2: ......

Страница 3: ...are 8 4 6 Preset Tare 8 5 0 RS 232 Communication 9 5 1 Pushbutton Keypad Print 9 5 2 USB Connection 9 6 0 Troubleshooting Testing 12 7 0 Maintenance 13 7 1 Basic Maintenance 13 7 2 Cleaning 13 8 0 Spe...

Страница 4: ...ii Contents Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webinars...

Страница 5: ...s displayed until reweighing is performed or until the scale zeros out This manual can be viewed and downloaded from the Rice Lake Weighing Systems website at www ricelake com Warranty information can...

Страница 6: ...the scale while the infant is in the weighing tray For accurate weighing the scale must be placed on a flat stable surface For accurate weighing verify proper operation according to the procedure desc...

Страница 7: ...ale select a sturdy flat surface on which to place the scale 3 3 1 Inserting Batteries The six AA Lithium Ion batteries that come with the scale offer an average of 25 hours of continuous use To insta...

Страница 8: ...ighing tray to the base assembly Figure 3 3 Attach Weighing Tray to Scale Base 3 3 2 Optional AC Power Supply The RL 650 Series neonatal scale has an optional 120VAC or 230VAC adapter to use when powe...

Страница 9: ...cale press the Tare key to tare out the scale and return it to zero The display annunciator turns to Net Zero This key zeros the weight of the scale It attempts to zero for three seconds and works onl...

Страница 10: ...es that the scale weight has been locked in and will not disappear from the scale display 5 Unit length This annunicator displays the unit s length either in inches or centimeters 6 Unit weight This a...

Страница 11: ...rvice provider 4 3 Weighing Use the following steps to weigh an infant or neonate on the RL 650 Series neonatal scale 1 Press the On Off key to turn the scale on 2 Wait for the Zero to appear in the d...

Страница 12: ...still on Clear the scale of all items and the weight of the extra load remains stored in memory 5 To cancel the tare weight press the Tare key until Net disappears from the display The display turns b...

Страница 13: ...as the weight was either in motion or out of range The print will show on the display when it is transmitting data If displaying weight the scale will send out the following 21 character string xxxxxx...

Страница 14: ...re 5 2 Download Screen for Hardware Wizard The screen prompts listed below show how to navigate through the screens 2 Select No not this time and then select Next 3 Select Install the software automat...

Страница 15: ...ndicator to be able to see the information being sent to the PC A port needs to be established so select the port that is assigned to Parpar and print the ticket The following example tickets will pri...

Страница 16: ...ing object from the weighing tray from the scale Display did not show 0 0 before weighing Remove the infant zero the scale and begin the weighing process again Scale is not placed on a level surface P...

Страница 17: ...or basic maintenance 1 Check the overall appearance of the entire scale for any obvious signs of damage abuse etc 2 Inspect the condition of the optional AC adapter for cord cracking or fraying or for...

Страница 18: ...ptional Optional adapter is UL and C UL approved Battery Type AA Lithium Ion disposable Nominal Voltage 1 5 V Rated capacity 2900 mAh Working voltage 1 30V 200 mA discharge Environmental Operating tem...

Страница 19: ......

Страница 20: ...Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 www ricelake com www ricelake mx www ricelake eu www ricelake co in m ricelake com Rice Lake Weighing Systems is an ISO 9001 regi...

Отзывы: