Rice Lake RL-350-5 Скачать руководство пользователя страница 15

Operation

© 

Rice Lake Weighing Systems

 

 All Rights Reserved

11

3.7 Troubleshooting

Refer to the following table to check and correct any failure before contacting service personnel.

Symptom

Possible Cause

Corrective Action

Scale does not turn on

Dead batteries

Replace batteries or connect to AC power

Faulty electrical outlet

Use a different electrical outlet

Bad power supply

Replace adapter

Questionable weight or the scale does not 

zero

External object is interfering with the scale Remove the interfering object from the scale
Display did not show 0.0 before weighing Help the patient off the scale, zero the scale and begin the 

weighing process again

Scale is not placed on a level floor

Ensure scale is level and begin the weighing process again

Scale is out of calibration

Check the weight with a certified calibration weight 

Scale base is touching floor during a 

weighment

Adjust height of feet so fingers can slide between the base 

of scale and the floor all the way around the platform

The display shows a 

STOP

 message

The load on the scale exceeds the 

capacity of the scale

Remove the excess weight and use the scale according to 

manufacture specifications

The display shows 

LO Bat

 message

The battery is low

Replace batteries

The display shows E and Err messages as detailed below
E06

Identifier - ADC

AD too high

E07

AD too low

E10

Overload

Scale has been overloaded. Remove load from scale

E4L

BAT

Battery low, but still usable- one bar left on indicator display

E4U

Battery low and unstable - no bars left on indicator display

E11

CAL

Calibration Error - recalibrate scale

Err 1

Load cell cable may be plugged into wrong 

connection port

Ensure cable is connected to the load cell connection port.

Note: 

Load cell connection point is located underneath the 

curved plastic cover of the indicator. Remove four back 

retaining screws, remove curved back cover to access load 

cell connection point.

Err 2

Low saturation state (low A/D)

The load cell is not connected properly; Check the cables 

and mechanical connections; if the problem persists, replace 

the set of load cells

Err 3

High saturation state (high A/D)

See Err 2

Err 6

Unstable weight; Cannot calibrate

Check the load cell mechanical surroundings and ensure 

nothing is contacting the load cell and that the cables are 

properly welded

Err 7

Scale isn’t moving

Make sure feet are installed on the scale. Turn the feet all 

the way in and then back them out three full turns, then level 

the scale

SAT

Damaged load cell cable

Replace load cell cable

Load cell cable may be plugged into wrong 

connection port

Ensure cable is connected to the load cell connection port.

Note: 

Load cell connection point is located underneath the 

curved plastic cover of the indicator. Remove four back 

retaining screws, remove curved back cover to access load 

cell connection point.

Table 3-2.   Troubleshooting Table

Содержание RL-350-5

Страница 1: ...RL 350 Series Portable Bariatric Wheelchair Scales Software Version 11525 Technical Manual RL 350 5 RL 350 6 PN 163357 Rev B August 18 2022 ...

Страница 2: ... respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserves the right to make changes to the technology features specifications and design of the equipment without notice The most current version of this publication software firmware and all other product updates can be ...

Страница 3: ...ning or obtained by calling 715 234 9171 and asking for the training department Revision History This section tracks and describes the current and previous manual revisions for awareness of major updates and when the updates took place Revision Date Description B August 18 2022 Established a revision history formatted content to match other medical manuals software version 11525 Table i Revision L...

Страница 4: ...g 9 3 3 Hold Release Function 9 3 4 Preset Tare 9 3 5 Toggle Tare 9 3 6 Using the Body Mass Index BMI Function 10 3 6 1 LB Mode 10 3 6 2 KG Mode 10 3 7 Troubleshooting 11 4 0 Configuration 12 4 1 Setup Switch 12 4 2 General Navigation 12 4 2 1 Change Parameters 12 4 2 2 Enter Numbers 12 4 2 3 Save and Return to Main Menu 12 4 3 Configuration Mode 13 4 4 Programming Mode Menu 14 4 5 Default Menu 17...

Страница 5: ...s and has an optional AC adapter available The scale is set up to use motion sensing technology to determine actual weight of a moving patient The weight can be displayed in either pounds or kilograms and a tare weight can be entered Section 3 0 on page 8 of this manual explains the scale operation and how to obtain a tare weight Manuals resources and warranties are available from Rice Lake Weighi...

Страница 6: ...rt the scale while someone is on the scale Do not allow minors children or inexperienced persons to operate this scale Do not use in the presence of flammable materials Do not use this product if any of the components are loose or cracked Do not use near water Do not use the scale on slippery surfaces such as a wet floor Do not use this scale when a person s body or feet are wet such as after taki...

Страница 7: ...Parts box which contains the indicator display 2 1 1 Repacking Retain the packaging for use in the event that the scale must be returned or moved The product must be properly packed with sufficient packing materials Whenever possible use the original carton and packing materials when shipping the scale back Damage caused by improper packaging is not covered by the warranty 2 2 Scale Setup Use the ...

Страница 8: ...d to be made in order for the scale to sit properly on the floor 1 There must be adequate clearance between the scale base and the floor so screw each foot out counter clockwise two full turns This will ensure that there is enough clearance between the scale base and the floor Figure 2 3 Backing out the Scale Feet 2 Carefully unwrap the load cell cable located on the underside of the scale It will...

Страница 9: ...uce inaccurate weight readings 2 3 Cable Connections Ten feet of 4 wire cable to connect the scale to the weight indicator is supplied with each scale The cable is pre installed from the factory The cable must be routed to the indicator in a manner that will protect the cable from damage Cabling should exit the underside of the scale on the portable wheel side of the scale and not the ramp side s ...

Страница 10: ... to gain access to the load cell connection Figure 2 6 Remove Tilt Stand 2 Remove the four back retaining screws to remove the back cover to the indicator Figure 2 7 Remove Back Cover 3 Plug the end of the load cell cable into the load cell connection When it clicks the load cell cable is properly seated into the connection Figure 2 8 Load Cell Connection Point 4 Reinstall the back cover and attac...

Страница 11: ...be used for an extended period of time If the LO BAT indicator activates for accurate weighing replace the batteries or connect the scale to an AC power source as soon as possible 2 4 2 Power Connection An optional AC power adapter can be used when a power outlet is available Only use power adapters supplied by or purchased from Rice Lake Weighing Systems The use of a power adapter not from Rice L...

Страница 12: ...I Pressing the BMI key enables access to the BMI Body Mass Index mode defaults when scale is turned on The patient is gets on the scale weight stabilizes and press the BMI key The display then asks for the patient height to calculate out the patient BMI TARE Used to remove the weight initially of anything on the scale that shouldn t be included in the total weight of the patient on the scale CLEAR...

Страница 13: ...by the patient 1 Press to turn on the scale 0 0 appears on the display along with the ZERO annunciator 2 Place additional item s on the scale 3 Press until the display returns to 0 0 and NET annunciator appears on the display 4 Remove additional item s from the scale The weight displays with a negative symbol to the left of it Not removing the additional item s prior to patient weighing will also ...

Страница 14: ... adjust the height value 5 Press 6 The BMI value and BMI annunciator are shown on the display Press to return to the Weighing mode and the BMI function will be turned off 3 6 2 KG Mode 1 Ensure that the scale is at zero 2 Place the patient on the scale to obtain a weight The LOCK annunciator appears on the display 3 Press The BMI and CM annunciators appear on the display and a default height value...

Страница 15: ... high E07 AD too low E10 Overload Scale has been overloaded Remove load from scale E4L BAT Battery low but still usable one bar left on indicator display E4U Battery low and unstable no bars left on indicator display E11 CAL Calibration Error recalibrate scale Err 1 Load cell cable may be plugged into wrong connection port Ensure cable is connected to the load cell connection port Note Load cell c...

Страница 16: ...the menus and or parameters Press to enter a displayed menu and or parameter Press or to scroll through values Press to save the displayed selection and move to the next parameter 4 2 2 Enter Numbers Press to enter parameter value Press or to increment decrement numbers Press to move to the next digit Press to save value and move to the next parameter 4 2 3 Save and Return to Main Menu When a para...

Страница 17: ...ocated in the back of the scale to enter the setup parameters for the scale Use a small paper clip small screwdriver or other similar object to press the setup switch 4 Once the setup switch is pressed PROG displays 5 Scale can be configured using a series of menus accessed through the front panel when the scale is in Setup mode Figure 4 2 Top Level Menu 6 Press to advance to the desired menu 7 Pr...

Страница 18: ...Use the Up Down arrow keys to increment decrement numbers 7 When done press ENTER key to move to the next parameter LOAD LOAD This is the amount of weight applied during calibration Can also be changed in the calibration menu Value 200 lb The display toggles between a numeric value and LOAD If you do not want to change this value press the BMI key to move to the next setting Example from LOAD to A...

Страница 19: ... 10 20 50 100 Values in lb 1 2 5 10 20 50 100 200 0 2 0 5 1 0 2 0 5 0 10 0 20 0 0 1 The display toggles between a numeric value and ROUND The decimal point location is set to the DISP parameter display decimal point location If you do not want to change this value press the BMI key to move to the next setting Example from ROUND to DISP If you want to change the value use the following steps 1 Pres...

Страница 20: ... use the following steps 1 Press the ENTER key to change values 2 Use the Up Down arrow keys to change the available parameters 3 When done press ENTER key to move to the next parameter UNITS UNITS Units Selects the unit of measure It can be either Kg Lb Kg only or Lb only KG LB KG LB The display toggles between unit of measurements and UNIT If you do not want to change this value press the BMI ke...

Страница 21: ...he scale to enter the setup parameters for the scale Use a small paper clip small screwdriver or other similar object to press the setup switch on the back of the indicator under the cover 4 Prog displays Press to toggle along the parameter menu 5 Calib displays and enter the calibration parameters 6 Press and a numeric value is displayed which represents the amount of weight that is used for cali...

Страница 22: ...r on the scale until ID flashes 2 Press again 3 Continue to press to scroll through the various menu items 4 Once complete press again then Done displays 5 Press to start the weighing process Figure 4 5 Test Menu Parameter Choice Description RID Value Displays internal ID number VER Value Displays the current software version BAT Value Displays the current battery level VALUE Value Displays the ac...

Страница 23: ...e weight was either in motion or out of range If displaying weight and not BMI the scale will send out the following 21 character string xxxxxxxxx SP uu SP mmmmm SP CR LF Example 60 1 KG PATIENT SP WEIGHT SP 60 1 SP KG SP CR LF In BMI mode displaying the BMI value the scale will send out the following data PATIENT WEIGHT 60 1 KG PATIENT HEIGHT170 0 CM PATIENT BMI 20 8 Example in KG PATIENT SP WEIG...

Страница 24: ...pe Protocol Commands Name ESC Character ESC Value with Parameters Description Reading R R Tells PC the scale is sending a reading immediately following this is the value that is sent Example ESC R ESC W0200 0 ESC Nm ESC E Weight W Wnnn n The patient weight Example W02000 means 200 0 If scale is overloaded or under loaded 999 99 is returned Height H Hnnn n Patient height BMI B Bnn n Patient BMI Uni...

Страница 25: ...Z Zero the scale F Show flash values used for the first flash process L USB On Off not available on USB communication Table 5 5 Maintenance Protocol Commands Escape Protocol Example 1 Escape Protocol Example 2 ESC R ESC E Reading request EOP SCALE PC EOP Unit of Measure Metric ESC R ESC W0210 9 ESC Nm ESC E Reading Command Weight 0210 9 SCALE PC direction of communication Diagnose Battery Request ...

Страница 26: ...ing the use of the USB driver Driver protections may need to be temporarily disabled on Windows 10 or later computers to allow for the installation of the USB driver 1 Connect the scale s indicator to a PC using a USB Type B to USB Type A cable not included 2 Turn the indicator on In most cases the PC should find the driver and automatically configure the driver when the scale is plugged into a US...

Страница 27: ...sonnel Go through the following steps for basic maintenance Check the overall appearance of the entire scale for any obvious signs of damage Inspect the condition of the AC power adapter cord for cracking fraying or for broken or bent prongs 6 2 Cleaning Proper care and cleaning is essential to ensure a long life of accurate and effective operation Before beginning the cleaning process disconnect ...

Страница 28: ...to another Use the following steps to transport the scale or to store it 1 Carefully pick up the scale by standing on the side opposite of the wheels where the cabling comes out from the bottom of the scale platform Figure 6 1 Pick up the Scale Using Handles 2 With the scale upright wrap the cable around brackets on the underside of the scale and hang indicator on bracket 3 The scale can now be ro...

Страница 29: ... be configured Environmental Operating Temperature 50 F to 104 F 14 C to 40 C Storage Temperature 32 F to 158 F 0 C to 70 C Humidity 85 relative humidity Load Cell Excitation Rated Excitation 10 VDC Maximum Excitation 15 VDC Grade Level Requirements The supporting surface for the four feet of the scale must be level within 1 4 inch of horizontal Nominal Scale Height 3 0 inch 76 mm Certifications a...

Страница 30: ...RL 350 Series Wheelchair Scales 26 Visit our website www RiceLake com ...

Страница 31: ......

Страница 32: ...I 54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems Specifications subject to change without notice www ricelake com PN 163357 Rev B August 18 2022 ...

Отзывы: