Tracer AV Installation Manual

6. With 

WVAL

 displayed, press 

Zero

   to show

t h e   st o r e d   c a l i b r a t i o n   w e i g h t .   U s e   t h e
procedure shown in Figure 3-3  12 to enter the
actual value of the test weights to be used for
the span calibration. Press 

Zero

  

again 

to

store the entered 

WVAL

 value and go to span

calibration (

WSPAN

).

7. With 

WSPAN

 displayed, place test weights on

the scale and press 

Zero

   to calibrate span.

The indicator displays 

*CAL*

 while calibration

is in progress. When complete, the A/D count
for the span calibration is displayed. Press

Zero

   again to store the span calibration

value.

4.2

Standard Calibration

To calibrate the 

Tracer AV

 using the 

Zero

 and 

Units

keys, do the following: 

1. Place the 

Tracer AV

 in setup mode

To enter setup mode, remove the setup switch
access screw on the 

Tracer AV

 (See Figure 2-3

4). Once the screw is removed, use a small
screwdriver or pen tip to press the setup
switch. The display changes to show the word

DEFCAL

.

2. Remove all weight from the scale platform. 

If your test weights require hooks or other
items, place them on the scale for zero
calibration. 

3. Press the 

Units

   key until the display reads

CALIBR

 (see Figure 4-1). Press the 

Zero

  

key

to go to zero calibration (

WZERO

). 

4. With 

WZERO

 displayed, press 

Zero

  

to

calibrate zero. The indicator displays 

*CAL*

while calibration is in progress. When
c o m p l e t e ,   t h e   A / D   c o u n t   f o r   t h e   z e r o
calibration is displayed. 

DO NOT adjust this

value after WSPAN has been set!

 Press 

Zero

 

again to save the zero calibration value and go
to the next prompt (

WVAL

).

5. With 

WVAL

 displayed, press 

Zero

   to show

t h e   st o r e d   c a l i b r a t i o n   w e i g h t .   U s e   t h e
procedure shown in Figure 3-3  12 to enter the
actual value of the test weights to be used for
the span calibration. Press 

Zero

  

again 

to

store the entered 

WVAL

 value and go to span

calibration (

WSPAN

).

6. With 

WSPAN

 displayed, place test weights on

the scale and press 

Zero

   to calibrate span.

The indicator displays 

*CAL*

 while calibration

is in progress. When complete, the A/D count
for the span calibration is displayed. Press

Zero

   again to store the span calibration

value and go to the next prompt (

REZERO

).

7. The rezero function is used to remove a

calibration offset when hooks or other items
are used to hang the test weights. 

If no other apparatus was used to hang the
test weights during calibration, remove
the test weights and press 

Units

  

to

return to the CALIBR menu. 

If hooks or other items were used during
calibration, remove these and the test
weights from the scale. With all weight
removed, press 

Zero

   to rezero the scale.

This function adjusts the zero and span
calibration values. The indicator displays

*CAL*

 while the zero and span calibrations

are adjusted. When complete, the adjusted
A/D count for the zero calibration is
displayed. Press 

Zero

   to save the value,

t h e n   p r e s s  

U n i t s

    t o   r e t u r n   t o   t h e

CALIBR menu.

8. Press 

Units

   until the display reads 

CONFIG

,

and the display will exit the setup mode.

4.3

EDP Command Calibration

To calibrate the 

Tracer AV

 using EDP commands, the

EDP port must be connected to a terminal or personal
computer. See Section 2.3.3 on page 7 for EDP port
pin assignments; see Section 5.0 on page 26 for more
information about using EDP commands.

Once the 

Tracer AV

 is connected to the sending device,

do the following:

1. Place the 

Tracer AV

 in setup mode (display

reads 

CONFIG

) and remove all weight from the

scale platform. If your test weights require
hooks or other items, place them on the scale
for zero calibration. 

2. Send the WZERO EDP command to calibrate

zero. The indicator displays 

*CAL*

 while

calibration is in progress.

3. Place test weights on the scale and use the

WVAL command to enter the test weight
value in the following format:

 WVAL=

nnnnnn

<CR> 

4. Send the WSPAN EDP command to calibrate

span. The indicator displays 

*CAL*

 while

calibration is in progress.

5. To remove an offset value, clear all weight

from the scale, including hooks or other items
used to hang test weights, then send the
REZERO EDP command. The indicator
displays 

* C A L *

 while the zero and span

calibrations are adjusted. 

6. Send the KUPARROW EDP command to exit

setup mode. 

Содержание REVOLUTION Tracer AV

Страница 1: ...98316 Tracer AV Aviation Baggage Instrumentation Installation Manual To be the best by every measure...

Страница 2: ......

Страница 3: ...2 3 Cable Connections 4 2 3 1 Illustrations 5 2 3 2 Load Cells 7 2 3 3 Serial Communications 7 2 3 4 Digital I O 8 2 4 Board Removal 9 2 5 Replacement Parts 9 3 0 Configuration 10 3 1 Configuration M...

Страница 4: ...ta to a Personal Computer 31 5 2 2 Downloading Configuration Data from PC to Tracer AV 31 6 0 Appendix 32 6 1 Error Messages 32 6 1 1 Displayed Error Messages 32 6 1 2 Using the XE EDP Command 33 6 2...

Страница 5: ...ndary function for the setup mode The Tracer AV features include Four wire load cell connections remote sense connections are not provided The electronic data processing EDP port is full duplex RS 232...

Страница 6: ...isplayed weight is a gross or net weight Center of zero Gross weight is within 0 25 graduations of zero This annunciator lights when the scale is zeroed Standstill Scale is at standstill or within the...

Страница 7: ...Tracer AV to ensure all components are included and undamaged The shipping carton should contain the Tracer AV enclosure remote display remote keypad this manual two cables and a power cord If any par...

Страница 8: ...the power cord one to accommodate the load cell communications remote displays and keypads Figure 2 3 shows the assignments for the Tracer AV connections Details on pin assignments are shown in Table...

Страница 9: ...VS1 TVS5 TVS4 3 3 R7 R4 C16 C4 DIGIN 1 J1 LOAD CELL U2 D1 U19 U6 U9 SERIAL 1 9 1 1 8 1 1 Gn d 20mA Gnd TxD Gnd Gn d DI 2 EXC E XC SEN SEN SIG 5 1 1 1 20mA RxD RxD TxD DI 1 SIG Micro processo r DISPLA...

Страница 10: ...6 Tracer AV Installation Manual Figure 2 6 Remote Display Cutout Figure 2 7 Remote Display Drawing...

Страница 11: ...allations so leave sense jumper JP1 and JP2 on NOTE The DB 9 is factory pinned for four wire connections Leave sense jumpers JP1 and JP2 on 2 3 3 Serial Communications The pinned DB 9 connector marked...

Страница 12: ...acer AV the setpoint value will cause the display to flash Outputs are designed to sink not source switching current Each output is a normally open connector circuit capable of sinking 250 mA when act...

Страница 13: ...d cell cable J2 digital inputs J3 EDP J4 serial J6 digital outputs and J2 display See Figure 2 4 5 for connector locations 6 Remove the five screws from the CPU board then lift the board out of the en...

Страница 14: ...ty allows configuration data to be retrieved from one Tracer AV edited then downloaded to another To use Revolution do the following 1 Install the Revolution module on a PC running Windows 98 or later...

Страница 15: ...so display and edit zero calibration a d count value test weight value and span calibration a d count value CONFIG Configuration Configure grads zero tracking zero range motion band overload tare func...

Страница 16: ...To save page space menu choices are shown in vertical columns The factory default setting appears at the top of each column in bold type Most menu diagrams are accompanied by a table that describes al...

Страница 17: ...ions Maximum legal value varies depending on local regulations NOTE For scales using linear calibration do not set the zero tracking band to a value greater than that specified for the first lineariza...

Страница 18: ...immediate area of the scale Choices indicate the number of A D conversions that are averaged to obtain the displayed reading A higher number gives a more accurate display by minimizing the effect of a...

Страница 19: ...Level 3 submenu parameter descriptions DSPRAT 250MS 500MS 750MS 1SEC 1 5SEC 2SEC 2 5SEC 3SEC 4SEC 6SEC 8SEC Display rate Sets the update rate for displayed values Values are in milliseconds MS or sec...

Страница 20: ...88 8888 88 Decimal point location Determines the location of the decimal point or dummy zeros in the secondary unit display DSPDIV 5D 1D 2D Display divisions Selects the value of minimum division size...

Страница 21: ...t weight value WSPAN Display and edit the span calibration A D count value REZERO Press Zero to remove an offset value from the zero and span calibrations Use this parameter only after WZERO and WSPAN...

Страница 22: ...from the EDP or printer port TERMIN CR LF CR Termination character Selects termination character for data sent from the EDP or printer port EOLDLY 000 number End of line delay Sets the delay period in...

Страница 23: ...ht is at no load NONE allows tares to be cleared at any weight value NTEP and OIML modes allow a new tare to be acquired even if a tare is already present In CANADA mode the previous tare must be clea...

Страница 24: ...display or print the date DATSEP SLASH DASH SEMI Specifies the date separator character TIMFMT 24HOUR 12HOUR Specifies the format used to display or print the time TIMSEP COLAN COMMA Specifies the tim...

Страница 25: ...sed VALUE number Displays and edits the setpoint value TRIP HIGHER LOWER INBAND OUTBND Trips the setpoint when the weight is higher or lower than the setpoint value or is within or outside of the band...

Страница 26: ...is used to check the software version and reg version installed in the Tracer AV You can also check the model Figure 3 11 Version Menu XXXXXXX ALGOUT XXXXXXX XXXXXXX VERS DIGIN SETPNT PROGRM PFORMT SE...

Страница 27: ...ss screw on the Tracer AV See Figure 2 3 4 Once the screw is removed use a small screwdriver to press the setup switch The display changes to show the word DEFCAL 2 Remove all weight from the scale pl...

Страница 28: ...progress When complete the A D count for the span calibration is displayed Press Zero again to store the span calibration value and go to the next prompt REZERO 7 The rezero function is used to remov...

Страница 29: ...ings 4 5 More About Calibration The following topics provide additional information about compensating for environmental factors Section 4 5 1 and diagnostic information for determining expected zero...

Страница 30: ...simulate pressing the keys on the 2 button keypad Most commands can be used in both setup and weighing mode Several of the commands serve as pseudo keys providing functions that are not represented by...

Страница 31: ...se no spaces before or after the equal sign If you type an incorrect command or value the display reads Changes to the parameters are saved as they are entered but typically do not take effect until y...

Страница 32: ...n REZERO Rezero LC CD Set deadload coefficient value LC CW Set span coefficient value Table 5 5 CALIBR EDP Commands Command Description Values EDP BAUD EDP port baud rate 300 600 1200 2400 4800 9600 1...

Страница 33: ...nt kind GROSS NET SP1 VALUE SP2 VALUE Setpoint value number SP1 TRIP SP2 TRIP Trip Higher Lower Inband Outbnd SP1 BNDVAL SP2 BNDVAL Band value number SP1 HYSTER SP2 HYSTER Hysteresis number SP1 ACCESS...

Страница 34: ...time hhmm enter using 24 hour format SX Start EDP streaming OK or EX Stop EDP streaming OK or RS Reset system XG Transmit gross weight in displayed units nnnnnn UU where nnnnnn is the weight value UU...

Страница 35: ...eters to the PC as ASCII formatted text 5 2 2 Downloading Configuration Data from PC to Tracer AV Configuration data saved on a PC or floppy disk can be downloaded from the PC to a Tracer AV This proc...

Страница 36: ...D converter requires recalibration Call RLWS Service EFCKSM Printer format checksum error Call RLWS Service ELCKSM Load cell calibration checksum error Recalibrate load cells EIDATA Internal RAM chec...

Страница 37: ...mand can be used to remotely query which annunciators are currently displayed The ZZ command returns the currently displayed weight and a decimal number representing the LED annunciators currently lit...

Страница 38: ...Text or 100DMText Figure 6 1 Continuous Output Data Format STX POL wwwwwww UNIT G N S TERM ASCII 02 decimal Weight data 7 digits right justi ed with decimal point leading zero supression Overload Unde...

Страница 39: ...anel Display Characters Figure 6 2 shows the 7 segment LED character set used to display alphanumeric characters on the Tracer AV front panel Figure 6 2 Tracer AV Display Characters 9 E Q F R G S 0 H...

Страница 40: ...pounds and the secondary decimal point is set to 8888 88 the indicator will overflow if 5 tons or more are applied to the scale With 5 tons applied and a conversion factor of 2000 the secondary unit...

Страница 41: ...ltering effect is 20 4 8 8 With this configuration each A D reading has a 1 in 20 effect on the displayed weight value Setting the filters to 1 effectively disables digital filtering 6 6 2 DFSENS and...

Страница 42: ...casional spikes to 75 lb record 50 lb as the threshold weight value 3 Place the Tracer AV in setup mode and set the digital filters DIGFLx to eliminate the vibration effects on the scale Leave DFTHRH...

Страница 43: ...three seconds the test mode display automatically shifts to the first test menu function A DTST Figure 6 5 shows the Test Menu structure Table 6 5 summarizes the test menu functions Figure 6 5 Test Mo...

Страница 44: ...emory Test Green Pass Red Fail LED ANA Red Not Implemented Analog Option Card Digital Input Output Status DO 1 Digital Out 1 On when lit active low DO 2 Digital Out 2 On when lit active low DI 1 Digit...

Страница 45: ...Input Overload 12 V continuous static discharge protected RFI Protection Signal excitation and sense lines protected by capacitor bypass Analog Output Optional fully isolated voltage or current outpu...

Страница 46: ...ot caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all alleged non conformities Such equipment has not been modif...

Страница 47: ......

Страница 48: ...PN 98316 02 11...

Отзывы: