33
1.1.6
Normal Mode Commands
The serial transmit weight data commands (see Table 1-12) transmit data to the EDP port on demand. The SX and
EX commands are valid only in normal operating mode; all other commands are valid in either setup or normal
mode.
Command
Description
Values
SOURCE
Analog output source
GROSS, NET
OFFSET
Zero offset
0%, 20%
ERRACT
Error action
FULLSC, HOLD, ZEROSC
MIN
Minimum value tracked
0–999 999
MINNEG
Minimum negative
OFF, ON
MAX
Maximum value tracked
0–999 999
MAXNEG
Maximum negative
OFF, ON
TWZERO
Zero calibration
0–16 383
TWSPAN
Span calibration
0–16 383
Table 1-11. ALGOUT EDP Commands
Command
Description
Response Format
UID
Set unit ID
nnnnnnn
SD
Set date
MMDDYY, DDMMYY, YYMMDD (enter using DATEFMT
specified)
ST
Set time
hhmm (enter using 24-hour format)
SX
Start EDP streaming
OK
or
??
EX
Stop EDP streaming
OK
or
??
RS
Reset system
—
XA
Transmit accumulator value
nnnnnn UU
where
nnnnnn
is the weight value,
UU
is the units.
XG
Transmit gross weight in displayed units
XN
Transmit net weight in displayed units
XT
Transmit tare weight in displayed units
XG2
Transmit gross weight in non-displayed units
XN2
Transmit net weight in non-displayed units
XT2
Transmit tare weight in non-displayed units
XE
Query system error conditions
nnnnn nnnnn
See Section 1.1.2 on page 39 for detailed information about
the XE command response format.
Table 1-12. Normal Mode Commands