Rice Lake CellScale MSI-9850 Скачать руководство пользователя страница 7

Installation

3

2.0

Installation

2.1

Ethernet 10/100 Base-T Connection

The Ethernet interface defaults to use Dynamic Host Configuration Protocol (DHCP) for obtaining an IP address. It 
is not necessary to be running a DCHP server to discover the scale device on network. The 

Scale Discovery Utility

 

uses multi-cast IP to discover and configure the device.

Once the device has an IP address, further configuration can be done via the web page interface if necessary.

The Ethernet connection can be made to the network via a hub or switch, or can go directly to a computer Ethernet 
port using a crossover Ethernet cable. For details, please reference the networking hardware manual, or contact the 
network administrator.

2.2

Cabling

1

2

A

B

12

28

C

3

4

D

E

F

G

Figure 2-1. Cable Assembly

Rice Lake Weighing Systems provides a field-installable mating connector

(

PN 144622

) that allows customers to interface with their own  Ethernet cabling. The 

connector accepts cables up to 8 mm.

Cable Assembly

See Figure 2-1.

1. Slide the pressure nut (item 1) and the housing (item 2) over  the cable (view 

A).

2. Strip the cable sheath over a length of approximately 1'' (28 mm). Trim the 

braided shield to a length of  approximately 1/2'' (12 mm) (view B). See 
Table 2-1 for the core assignment.

3. Guide the individual cores fully into the marked  chambers. Make sure to use 

the correct chambers and only one core per chamber (View C).

4. Contact the cores by pressing the termination block (item 3) to the plug insert 

(item 4) according to views D and E. The  termination block engages at the 
end stop.

5. Pull the housing (item 2) over the plug insert (item 4), hold the plug insert 

tight and screw the housing onto it (View F).

6. Push the pressure nut (item 1) on the housing. Tighten the  pressure nut firmly 

(View G).

Table 2-1. Core Assignment

Pin

TIA 568 B

Color

Function

1

White/Orange

TD+

2

Orange

TD-

3

White/Green

RD+

4

Green

RD-

Содержание CellScale MSI-9850

Страница 1: ...Ethernet 802 11b Operator Manual PN 161444 Rev A Ethernet 802 11b...

Страница 2: ...marks of their respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserv...

Страница 3: ...2 Cabling 3 2 3 802 11b Connection 4 2 4 Configuring Network Settings 4 2 5 Scale Discovery Utility 4 2 6 Web Interface 6 2 6 1 Wireless Settings 8 2 7 Installation without Supporting Network Infrast...

Страница 4: ...ii Ethernet Option Operator Manual Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webinars...

Страница 5: ...ent unless this manual has been read and all instructions are understood Failure to follow the instructions or heed the warnings could result in injury or death Contact any Rice Lake Weighing Systems...

Страница 6: ...ne scale 4 Configure the IP settings see the network administrator if information is unknown 5 Open the web interface and configure the network and security settings as necessary 6 Reboot the Ethernet...

Страница 7: ...e mating connector PN 144622 that allows customers to interface with their own Ethernet cabling The connector accepts cables up to 8 mm Cable Assembly See Figure 2 1 1 Slide the pressure nut item 1 an...

Страница 8: ...e Discovery Utility Note Scale Discovery Utility found on the Rice Lake website may need to be added to the security on the computer for the Scale Discovery Utility to run See Section 3 2 on page 10 T...

Страница 9: ...overy Utility Device Information 4 Press Configure A Device Configuration dialog box displays 5 Select to Automatically adjust the network settings via DHCP or Manually configure network settings Figu...

Страница 10: ...end of line string defaults to carriage return line feed Semicolon is a special command that instructs the scale to reset its command parser Note Use the semicolon command to reset the scale parser i...

Страница 11: ...ge 3 Select Network from the menu to configure the network interface The Network Configuration menu displays Figure 2 8 Web Interface Network Configuration 4 Enter the network settings and select Appl...

Страница 12: ...pported wireless security capabilities WEP Wired Equivalent Privacy 64 128 bit encryption RC4 WPA WPA2 802 11i Strong SSL 3 0 TLS 1 0 based encryption DES 56 bit 3DES 168 bit AES 128 256 bit 128 bit T...

Страница 13: ...rarily disable any Firewall running on the computer Start the program as a Windows Service or standalone depending on specific installation 2 Cycle power on the scale device 3 Run the Scale Discovery...

Страница 14: ...d Use the host commands to talk to the scale or observe the continuous print string if the crane scale is configured so 6 To end the session save the connection and close the terminal program 3 2 Secu...

Страница 15: ...amily These radios are not compatible with each other 3 4 Wireless Troubleshooting The main challenge is getting the 802 11b equipped crane scale to associate with a Wireless Access Point WAP Once thi...

Страница 16: ...nique MAC address The MAC address is on a sticker on the Digi device server The number is displayed as 12 hexadecimal digits usually starting with 00 40 9D Multi cast IP A method of forwarding IP data...

Страница 17: ......

Страница 18: ...WI 54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems is an ISO 9001 registered company Rice Lake Weighing Systems Spe...

Отзывы: