10

Ethernet 802.11b

3.0

Appendix 

3.1

Using a Terminal Program

Hyper Terminal and similar utilities are communications programs that can be used to connect to an Ethernet 
interface for communications and testing.

1. Start the terminal program.

2. Enter a name for the connection.

3. Configure the connection to use TCP/IP.

4. Enter the device address and port number 2101.

5. The connection is now established.  Use the host commands to talk to the scale or observe the continuous 

print string if the crane scale is configured so.

6. To end the session save the connection and close the terminal program.

3.2

Security Settings

Java security settings may need to be modified on the computer in order for the scale discovery utility to run.

1. Open the 

Control Panel

 panel on the computer.

2. Select 

Programs

 (for category view).

3. Select 

Java

 to open the 

Java Control Panel

.

4. Select the 

Security

 tab.

5. Click the 

Edit Site List…

 button.

6. Add the 

Scale Discovery Utility

see Figure 3-1

.

Figure 3-1. Add MSI Site to Security

Содержание CellScale MSI-9850

Страница 1: ...Ethernet 802 11b Operator Manual PN 161444 Rev A Ethernet 802 11b...

Страница 2: ...marks of their respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserv...

Страница 3: ...2 Cabling 3 2 3 802 11b Connection 4 2 4 Configuring Network Settings 4 2 5 Scale Discovery Utility 4 2 6 Web Interface 6 2 6 1 Wireless Settings 8 2 7 Installation without Supporting Network Infrast...

Страница 4: ...ii Ethernet Option Operator Manual Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webinars...

Страница 5: ...ent unless this manual has been read and all instructions are understood Failure to follow the instructions or heed the warnings could result in injury or death Contact any Rice Lake Weighing Systems...

Страница 6: ...ne scale 4 Configure the IP settings see the network administrator if information is unknown 5 Open the web interface and configure the network and security settings as necessary 6 Reboot the Ethernet...

Страница 7: ...e mating connector PN 144622 that allows customers to interface with their own Ethernet cabling The connector accepts cables up to 8 mm Cable Assembly See Figure 2 1 1 Slide the pressure nut item 1 an...

Страница 8: ...e Discovery Utility Note Scale Discovery Utility found on the Rice Lake website may need to be added to the security on the computer for the Scale Discovery Utility to run See Section 3 2 on page 10 T...

Страница 9: ...overy Utility Device Information 4 Press Configure A Device Configuration dialog box displays 5 Select to Automatically adjust the network settings via DHCP or Manually configure network settings Figu...

Страница 10: ...end of line string defaults to carriage return line feed Semicolon is a special command that instructs the scale to reset its command parser Note Use the semicolon command to reset the scale parser i...

Страница 11: ...ge 3 Select Network from the menu to configure the network interface The Network Configuration menu displays Figure 2 8 Web Interface Network Configuration 4 Enter the network settings and select Appl...

Страница 12: ...pported wireless security capabilities WEP Wired Equivalent Privacy 64 128 bit encryption RC4 WPA WPA2 802 11i Strong SSL 3 0 TLS 1 0 based encryption DES 56 bit 3DES 168 bit AES 128 256 bit 128 bit T...

Страница 13: ...rarily disable any Firewall running on the computer Start the program as a Windows Service or standalone depending on specific installation 2 Cycle power on the scale device 3 Run the Scale Discovery...

Страница 14: ...d Use the host commands to talk to the scale or observe the continuous print string if the crane scale is configured so 6 To end the session save the connection and close the terminal program 3 2 Secu...

Страница 15: ...amily These radios are not compatible with each other 3 4 Wireless Troubleshooting The main challenge is getting the 802 11b equipped crane scale to associate with a Wireless Access Point WAP Once thi...

Страница 16: ...nique MAC address The MAC address is on a sticker on the Digi device server The number is displayed as 12 hexadecimal digits usually starting with 00 40 9D Multi cast IP A method of forwarding IP data...

Страница 17: ......

Страница 18: ...WI 54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems is an ISO 9001 registered company Rice Lake Weighing Systems Spe...

Отзывы: